Angle alpha orientation and magnitude distribution in a cataract surgery population
Authors:
- 1 March 2020
- journal article
- conference paper
- Published by Wolters Kluwer Health in Journal of Cataract & Refractive Surgery
- Vol. 46 (3) , 372-377
Abstract
Purpose: To report normative angle alpha orientation and magnitude distribution in a cataract surgery patient population. Setting: U.S. academic medical center. Design: Retrospective review. Methods: A total of 11 871 wavefront aberrometer/corneal topographer angle alpha measurements were performed over 40 months, yielding 8773 bilateral measurements (73.9%) with zero bad wavefront datapoints. The analysis was limited to the first right-eye scan for 3382 unique patients. The angle alpha magnitudes (millimeters) and orientations (degrees) were analyzed for these unique right eyes. Results: For the 3,382 unique patients, the mean angle alpha magnitude was 0.44 ± 0.15 mm (median, 0.44 mm; 25th and 75th percentiles 0.34 mm, 0.53 mm). Angle alpha orientation was predominantly horizontal (P < .01), with a mean of 186 ± 32 degrees. The expected point of intraocular lens (IOL) centration (EPIC) based on the geometric center of the corneal limbus was temporal to the visual axis in 3212 eyes (95%), nasal in 92 eyes (2.7%), inferior in 56 eyes (1.7%), and superior in 22 eyes (0.6%). The mean angle alpha magnitude was 0.3 mm or less in 607 eyes (18%) and 0.5 mm or greater in 1089 eyes (32%). Conclusions: Angle alpha is a predominantly horizontal phenomenon with a mean EPIC of 0.44 mm temporal to the visual axis. This information may assist in determining eligibility for patients in multifocal IOL implantation.Keywords
This publication has 14 references indexed in Scilit:
- Streamlined method for anchoring cataract surgery and intraocular lens centration on the patient’s visual axisJournal of Cataract & Refractive Surgery, 2018
- Special Report: American Academy of Ophthalmology Task Force Summary Statement for Measurement of Tilt, Decentration, and Chord LengthOphthalmology, 2017
- Optical Zone Centration Accuracy Using Corneal Fixation-based SMILE Compared to Eye Tracker-based Femtosecond Laser-assisted LASIK for MyopiaJournal of Refractive Surgery, 2015
- Centration axis in refractive surgeryEye and Vision, 2015
- The Subject-Fixated Coaxially Sighted Corneal Light Reflex: A Clinical Marker for Centration of Refractive Treatments and DevicesAmerican Journal of Ophthalmology, 2014
- Multifocal intraocular lenses: Relative indications and contraindications for implantationJournal of Cataract & Refractive Surgery, 2014
- Negative dysphotopsia: The enigmatic penumbraJournal of Cataract & Refractive Surgery, 2012
- Optical aberrations and alignment of the eye with ageJournal of Vision, 2010
- Effect of decentration and tilt on the image quality of aspheric intraocular lens designs in a model eyeJournal of Cataract & Refractive Surgery, 2009
- Aberrations induced in wavefront-guided laser refractive surgery due to shifts between natural and dilated pupil center locationsJournal of Cataract & Refractive Surgery, 2006