Thermoluminescence of LiF: dependence of thermal history
Author:
- 1 January 1969
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 14 (1) , 107-118
Abstract
The influence of the thermal history on the yield of thermoluminescence has been investigated. Dosimeters (mostly LiF-Teflon) have been stored at various temperatures for various times (pre-annealing) prior to irradiation and readout. Temperatures as high as 300 degrees C may give the LiF-crystal a state of thermal equilibrium in seconds, while the same equilibrium will not be reached for months at 80 degrees C. For precision dosimetry, pre-annealing at a high temperature should last only long enough to give thermal equilibrium and to empty all filled traps. Especially important is reproducibility of the cooling rate. With an appropriate heating circuit, the demands of both pre-annealing and cooling rate can be satisfied with the readout heating.Keywords
This publication has 6 references indexed in Scilit:
- Thermal Annealing Effects on the Thermoluminescence of LiFHealth Physics, 1966
- The Properties of Lithium Fluoride-Teflon Thermoluminescence Dosimeters and Their Use in RadiologyRadiology, 1966
- Präparation polykristalliner LiF-Speicherphosphore für die ThermolumineszenzdosimetrieIsotopenpraxis Isotopes in Environmental and Health Studies, 1966
- Thermoluminescent Radiation Dosimetry Utilizing LiFHealth Physics, 1964
- CaSO4(Mn) Thermoluminescence Dosimeters for Small Doses of Gamma RadiationReview of Scientific Instruments, 1962
- Effect of Impurities on the Flow Stress of LiF CrystalsJournal of Applied Physics, 1962