Detection of preanalytical errors in arterial blood gas analysis
Authors:
Open Access
- 15 June 2022
- journal article
- research article
- Published by Croatian Society for Medical Biochemistry and Laboratory Medicine in Biochemia Medica
- Vol. 32 (2) , 244-252
Abstract
Blood gas analysis (BGA) is an essential test used for years to provide vital information in critically ill patients. However, the instability of the blood gases is a problem. We aimed to evaluate time and temperature effects on blood gas stability. Arterial blood was collected from 20 patients into syringes. Following BGA for baseline, syringes were divided into groups to stand at 4°C and 22°C for 30, 60, 90, 120 minutes. All were tested for pH, partial pressure of carbon dioxide (pCO2), partial pressure of oxygen (pO2), oxygen saturation (sO2), oxyhemoglobin (O2Hb), sodium, potassium, glucose, lactate, oxygen tension at 50% hemoglobin saturation (p50), and bicarbonate. A subgroup analysis was performed to detect the effect of air on results during storage. Percentage deviations were calculated and compared against the preset quality specifications for allowable total error. At 4°C, pO2 was the least stable parameter. At 22°C, pO2 remained stable for 120 min, pH and glucose for 90 min, lactate and pCO2 for 60 min. Glucose and lactate were stable when chilled. Air bubbles interfered pO2 regardless of temperatures, whereas pCO2 increased significantly at 22°C after 30 min, and pH decreased after 90 min. Bicarbonate, sO2, O2Hb, sodium, and potassium were the unaffected parameters. Correct BGA results are essential, and arterial sample is precious. Therefore, if immediate analysis cannot be performed, up to one hour, syringes stored at room temperature will give reliable results when care is taken to minimize air within the blood gas specimen.Keywords
This publication has 24 references indexed in Scilit:
- Effect of Sample Storage Temperature and Time Delay on Blood Gases, Bicarbonate and pH in Human Arterial Blood SamplesIranian Red Crescent Medical Journal, 2015
- Effects of post-sampling analysis time, type of blood samples and collection tubes on values of blood gas testingMedicinski Glasnik, 2014
- Sepsis-associated hyperlactatemiaCritical Care, 2014
- Clinical biochemistry laboratory rejection rates due to various types of preanalytical errorsBiochemia Medica, 2014
- Different manufacturers of syringes: A new source of variability in blood gas, acid–base balance and related laboratory test?Clinical Biochemistry, 2012
- Agreement between paired blood gas values in samples transported either by a pneumatic system or by human courierClinical Chemistry and Laboratory Medicine (CCLM), 2011
- Comparison of Arterial and Venous Blood Gases and the Effects of Analysis Delay and Air Contamination on Arterial Samples in Patients with Chronic Obstructive Pulmonary Disease and Healthy ControlsRespiration, 2010
- Errors in measuring blood gases in the intensive care unit: Effect of delay in estimationJournal of Critical Care, 2003
- Current databases on biological variation: pros, cons and progressScandinavian Journal of Clinical and Laboratory Investigation, 1999
- Effects of sample storage time, temperature and syringe type on blood gas tensions in samples with high oxygen partial pressures.Thorax, 1994