Validation of a discovery MI 4-ring model according to the NEMA NU 2-2018 standards: from Monte Carlo simulations to clinical-like reconstructions
Open Access
- 31 January 2024
- journal article
- research article
- Published by Springer Nature in EJNMMI Physics
- Vol. 11 (1) , 1-23
Abstract
Background: We propose a comprehensive evaluation of a Discovery MI 4-ring (DMI) model, using a Monte Carlo simulator (GATE) and a clinical reconstruction software package (PET toolbox). The following performance characteristics were compared with actual measurements according to NEMA NU 2-2018 guidelines: system sensitivity, count losses and scatter fraction (SF), coincidence time resolution (CTR), spatial resolution (SR), and image quality (IQ). For SR and IQ tests, reconstruction of time-of-flight (TOF) simulated data was performed using the manufacturer’s reconstruction software. Results: Simulated prompt, random, true, scatter and noise equivalent count rates closely matched the experimental rates with maximum relative differences of 1.6%, 5.3%, 7.8%, 6.6%, and 16.5%, respectively, in a clinical range of less than 10 kBq/mL. A 3.6% maximum relative difference was found between experimental and simulated sensitivities. The simulated spatial resolution was better than the experimental one. Simulated image quality metrics were relatively close to the experimental results. Conclusions: The current model is able to reproduce the behaviour of the DMI count rates in the clinical range and generate clinical-like images with a reasonable match in terms of contrast and noise.Keywords
Funding Information
- GE Healthcare
This publication has 43 references indexed in Scilit:
- STIR: software for tomographic image reconstruction release 2Physics in Medicine & Biology, 2012
- Comparative evaluation of scatter correction in 3D PET using different scatter-level approximationsAnnals of Nuclear Medicine, 2011
- GATE V6: a major enhancement of the GATE simulation platform enabling modelling of CT and radiotherapyPhysics in Medicine & Biology, 2011
- Noise and signal properties in PSF-based fully 3D PET image reconstruction: an experimental evaluationPhysics in Medicine & Biology, 2010
- Monte Carlo simulations of the count rate performance of a clinical whole‐body MR/PET scannerMedical Physics, 2009
- Fifty years of Monte Carlo simulations for medical physicsPhysics in Medicine & Biology, 2006
- Validation of a Monte Carlo simulation of the Philips Allegro/GEMINI PET systems using GATEPhysics in Medicine & Biology, 2006
- Validation of GATE Monte Carlo simulations of the GE Advance/Discovery LS PET scannersMedical Physics, 2005
- The role of PET in monitoring therapyCancer Imaging, 2005
- Relevance of accurate Monte Carlo modeling in nuclear medical imagingMedical Physics, 1999