A Novel Technique for Identification of Doppler Microembolic Signals Based on the Coincidence Method
Authors:
- 1 April 1996
- journal article
- research article
- Published by Wolters Kluwer Health in Stroke
- Vol. 27 (4) , 683-686
Abstract
Background and Purpose The applicability of a novel differentiation technique in embolus detection based on the coincidence principle and using a multigate probe was evaluated in this study. Methods According to the coincidence method, high-intensity transients should only be classified as microembolic signals if they appear sequentially in the two sample volumes monitored and within a defined time window calculated from the blood velocity and the spatial distance between the insonation depths. Part A: microbubbles were introduced in a continuous-flow bench model of the middle cerebral artery to evaluate the accuracy of the multigate probe in embolus detection. Part B: in the subjects and patients, the minimal and maximum time delays in the appearance of microembolic signals in the two middle cerebral artery sample volumes were calculated as 0.01 second and set at 0.1 second, respectively. The multigate probe was used to monitor (1) 5 normal volunteers in whom 1008 artifact signals were produced, (2) 2 patients undergoing aortic valve replacement surgery, and (3) 12 patients with potential cardiac or carotid embolic sources. Results In the bench model, 95.5% of microembolic signals produced by microbubbles appeared in the two sample volumes with a time delay between 0.02 and 0.05 second, while in the remaining 4.5% a shorter passage time of 0.01 second was measured. A total of 1968 high-intensity signals were recorded in subjects and patients. All but 20 of these (99%) appeared in both monitoring channels within the above time frame. To summarize, 996 (98.8%) of the 1004 artifact signals and 943 (98.1%) of the 961 microembolic signals were correctly classified. Conclusions Application of the coincidence theory to distinguish microembolic signals from artifacts provides a promising new technique with high sensitivity and specificity that could decisively improve the validity of embolus detection.Keywords
This publication has 4 references indexed in Scilit:
- Real-time identification of cerebral microemboli with US feature detection by a neural network.Radiology, 1994
- Processing Doppler ultrasound signals from blood-borne emboliUltrasound in Medicine & Biology, 1994
- Computerized detection of cerebral emboli and discrimination from artifact using Doppler ultrasound.Stroke, 1993
- A Multigate Pulsed Doppler System with Serial Data ProcessingIEEE Transactions on Sonics and Ultrasonics, 1981