CMUT-in-CMOS ultrasonic transducer arrays with on-chip electronics
- 1 June 2009
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- No. 2159547X,p. 1222-1225
Abstract
Integration of a 2D ultrasound transducer array with front-end signal conditioning electronics poses significant fabrication and packaging challenges due to the required number of elements and small element pitch. In this paper, we demonstrate a CMUT-in-CMOS approach that produces ultrasonic transducers and circuits on a single chip using a standard foundry CMOS process, augmented only with two blanket post-process steps for sacrificial etching and cavity sealing. 2D arrays as large as 16 times 16, with integrated on-chip buffers, switches and multiplexers were fabricated and successfully characterized. This approach provides a straightforward path to commercially viable 2D arrays for 3D medical imaging.Keywords
This publication has 5 references indexed in Scilit:
- Integration of 2D CMUT arrays with front-end electronics for volumetric ultrasound imagingIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 2008
- Enhanced echographic images obtained improving the membrane structural layer of the cMUT probe2005
- Low temperature fabrication of immersion capacitive micromachined ultrasonic transducers on silicon and dielectric substratesIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 2004
- Volumetric ultrasound imaging using 2-D CMUT arraysIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 2003
- Micromachined ultrasound transducers with improved coupling factors from a CMOS compatible processUltrasonics, 2000