An accurate programmable ECG simulator
Authors:
- M. NasorMNM. Nasor
- 1 January 2001
- journal article
- Published by Taylor & Francis in Journal of Medical Engineering & Technology
- Vol. 25 (3) , 97-102
Abstract
This article reports the design and development of an ECG simulator intended for use in the testing, calibration and maintenance of electrocardiographic equipment. It generates a lead II signal having a profile that varies with heart rate in a manner which reflects the true in vivo variation. Facilities are provided for user adjustment of heart rate, signal amplitude, QRS complex up-slope, and the relative amplitudes of the P-wave and T-wave. The heart rate can be set within the range 30-200 beats min(-1) in steps of 1 beat min(-1). The amplitude of the QR5 complex can be adjusted from 0.1-20 mV in 0.1 mV steps, while its up-slope can be set between 10 and 50 ms with a 1 ms resolution. The amplitude of the P-wave can be varied from 5-40% and that of the T-wave from 10-80% of the amplitude of the QRS complex with a 1% resolution.Keywords
This publication has 7 references indexed in Scilit:
- The duration of the QT interval as a function of heart rate: A derivation based on physical principles and a comparison to measured valuesAmerican Heart Journal, 1985
- Signal Distortion in the Electrocardiogram Due to Inadequate Phase ResponseIEEE Transactions on Biomedical Engineering, 1983
- The relationship between QT interval and heart rate during physiological exercise and pacing.Japanese Heart Journal, 1981
- Electrocardiographic distortions caused by inadequate high-frequency response of direct-writing electrocardiographsAmerican Heart Journal, 1967
- The low-frequency response of electrocardiographs, a frequent source of recording errorsAmerican Heart Journal, 1966
- The normal Q-T intervalAmerican Heart Journal, 1962
- THE NORMAL DURATION OF THE ELECTROCARDIOGRAPHIC VENTRICULAR COMPLEXJournal of Clinical Investigation, 1936