Comparison of end-expiratory respiratory gating methods for PET/CT
Authors:
- Heidi HirviHHHeidi Hirvi
- Anna MäkipääAMAnna Mäkipää
- Jarmo TeuhoJTJarmo Teuho
- Tuomas KoivumäkiTKTuomas Koivumäki
- Marko M. MäkeläMMMarko M. Mäkelä
- Mika TeräsMTMika Teräs
- 24 June 2014
- journal article
- conference paper
- Published by Medical Journals Sweden AB in Acta Oncologica
- Vol. 53 (8) , 1079-1085
Abstract
Respiratory motion in positron emission tomography/computed tomography (PET/CT) causes underestimation of standardized uptake value (SUV) and variation of lesion volume, while PET and CT attenuation correction (CTAC) mismatch may introduce artefacts. The aim was to compare end-expiratory gating methods of PET and CTAC. Three methods named the minimum-constant, slope-based and amplitude-median were developed and evaluated on gating efficiency. Method evaluation and optimization was performed on 23 simulated and 23 recorded signals from a mixed patient group. The optimized methods were applied in PET/CT imaging of seven patients, consisting of non-gated CTAC, whole-body PET and four-dimensional (4D) PET/CT. Gating efficiency was evaluated by preservation of the respiratory signal, PET-CTAC alignment, image noise and measurement of lesion SUV maximum (SUVmax), SUV mean (SUVmean) and volume. The methods were evaluated with non-gated PET and end-expiratory phase of five-bin phase-gated PET. End-expiratory gated 4D-CTAC and averaged CTAC were compared for attenuation correction of end-expiratory gated PET. Mean fraction of data preserved was larger (23-34%) with end-expiratory gating compared to phase-gated PET. End-expiratory gating showed increased SUVmax (8.2-8.4 g/ml), SUVmean (5.7-5.8 g/ml) and decreased lesion volume (-11.3-16.8%) compared to non-gated PET (SUVmax 6.2 g/ml, SUVmean 4.7 g/ml) and phase-gated PET (SUVmax 8.0 g/ml, SUVmean 5.6 g/ml). Using averaged CTAC and end-expiratory 4D-CTAC produced similar results concerning SUVmax, with less than 5% difference. Additionally, CTAC-PET-mismatch was minimal when end-expiratory 4D-CTAC was used. End-expiratory gating in PET/CT results in SUVmax and SUVmean increase and reduced lesion volume compared to non-gated PET and phase-gated PET. End-expiratory 4D-CTAC or averaged CTAC will offer similar accuracy for attenuation correction of end-expiratory gated PET.Keywords
This publication has 19 references indexed in Scilit:
- The effect of breathing irregularities on quantitative accuracy of respiratory gated PET/CTMedical Physics, 2012
- Optimal gating compared to 3D and 4D PET reconstruction for characterization of lung tumoursEuropean Journal of Nuclear Medicine and Molecular Imaging, 2011
- The impact of respiratory motion on tumor quantification and delineation in static PET/CT imagingPhysics in Medicine & Biology, 2009
- Deep-Inspiration Breath-Hold PET/CT of Lung Cancer: Maximum Standardized Uptake Value Analysis of 108 PatientsJournal of Nuclear Medicine, 2008
- Evaluation of the combined effects of target size, respiratory motion and background activity on 3D and 4D PET/CT imagesPhysics in Medicine & Biology, 2008
- Respiratory gating in positron emission tomography: A quantitative comparison of different gating schemesMedical Physics, 2007
- Design of respiration averaged CT for attenuation correction of the PET data from PET/CTMedical Physics, 2007
- Attenuation correction of PET cardiac data with low‐dose average CT in PET/CTMedical Physics, 2006
- Effect of motion on tracer activity determination in CT attenuation corrected PET images: A lung phantom studyMedical Physics, 2005
- Four‐dimensional (4D) PET/CT imaging of the thoraxMedical Physics, 2004