Abstract
This paper addresses one of the major problems in interventional magnetic resonance imaging (MRI): the visualization of interventional devices. For visualization locally induced magnetic fields are used, which disturb the homogeneity of the main magnetic field of the MR scanner. This results in signal loss in the vicinity of the device due to intravoxel dephasing, and leads to a disturbance of the phase image. The local fields are established by a low current in a closed copper loop along the device. This method is introduced as a means for catheter visualization. The basic theory behind this method is presented. Simulations are performed to determine the effect of intravoxel dephasing, without interfering effects like susceptibility or radio-frequency artifacts. Scanned and simulated data is used to verify the theoretical consideration. Different configurations of wire loops are discussed and two types of catheter visualization scans are proposed. Results from a pig study show that this methods holds promise for intravascular interventions under MRI guidance.Keywords
This publication has 9 references indexed in Scilit:
- Catheter visualization using locally induced, actively controlled field inhomogeneitiesMagnetic Resonance in Medicine, 1997
- Catheter tracking using continuous radial MRIMagnetic Resonance in Medicine, 1997
- Distributed large-scale simulation of magnetic resonance imaging.Magnetic Resonance Materials in Physics, Biology and Medicine, 1997
- Visualization of dedicated catheters using fast scanning techniques with potential for MR‐guided vascular interventionsMagnetic Resonance in Medicine, 1996
- Biopsy needle susceptibility artifactsMagnetic Resonance in Medicine, 1996
- Verhalten des Flußkontrastes bei schneller MR-Angiographie mit einer EKG-getriggerten, schichtselektiven Inversion-recovery-Turbo-FLASH-TechnikRöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren, 1996
- Intravascular MR tracking catheter: preliminary experimental evaluation.American Journal of Roentgenology, 1995
- Nonsusceptibility artifacts due to metallic objects in MR imagingJournal of Magnetic Resonance Imaging, 1995
- Real‐time position monitoring of invasive devices using magnetic resonanceMagnetic Resonance in Medicine, 1993