Noninvasive measurements of skeletal muscle hemodynamics using frequency-domain near-infrared spectroscopy: contributions from adipose and bone tissues
Open Access
- 1 July 2025
- journal article
- general
- Published by SPIE-Intl Soc Optical Eng in Biophotonics Discovery
- Vol. 2 (03) , 1-21
Abstract
SignificanceNoninvasive optical assessments of skeletal muscle perfusion and metabolism are crucial for studying muscle physiology. However, measurements can be affected by contributions from overlying adipose and underlying bone tissues.AimWe aim to investigate the influence of bone tissue on frequency-domain near-infrared spectroscopy (FD-NIRS) measurements of the human forearm using different data types (intensity and phase), source–detector separations (25 and 37 mm), and acquisition modes (single-distance and dual-slope).ApproachFD-NIRS data were collected on the forearms of 14 healthy subjects using a custom optical probe designed to regulate probe-skin contact pressure. Tissue blood flow and oxygen consumption were measured with venous and arterial occlusion protocols, respectively. In vivo findings were compared with diffusion theory simulations for two- and three-layer media. Ultrasound imaging quantified adipose tissue thickness (1.5 to 9.5 mm) and bone depth (14 to 26 mm).ResultsExperimental results for bone depths below 20 mm could not be reproduced by two-layer simulations, whereas they aligned well with three-layer simulations mimicking adipose, muscle, and bone tissues. Adipose tissue and bone contributions depend on data type, measurement mode, and source–detector separation.ConclusionsBoth adipose and bone tissues may contribute to noninvasive FD-NIRS measurements of skeletal muscle, depending on tissue anatomy and data collection methods.Keywords
Funding Information
- National Institutes of Health (R01-EB029414)
This publication has 40 references indexed in Scilit:
- Light diffusion in a turbid cylinder II Layered caseOptics Express, 2010
- Quality Control, Accuracy, and Prediction Capacity of Dual Energy X-ray Absorptiometry Variables and Data AcquisitionJournal of Physiological Anthropology, 2008
- Investigation of depth selectivity of polarization gating for tissue characterizationOptics Express, 2005
- Adipose tissue thickness affects in vivo quantitative near-IR spectroscopy in human skeletal muscleClinical Science, 2001
- Influence of a fat layer on muscle oxygenation measurement using near-IR spectroscopy: quantitative analysis based on two-layered phantom experiments and Monte Carlo simulationFrontiers of Medical and Biological Engineering, 2000
- Blood flow and oxygen consumption with near-infrared spectroscopy and venous occlusion: spatial maps and the effect of time and pressure of inflationJournal of Biomedical Optics, 2000
- Near-infrared optical properties ofex vivohuman skin and subcutaneous tissues measured using the Monte Carlo inversion techniquePhysics in Medicine & Biology, 1998
- Absorption spectroscopy in tissue-simulating materials: a theoretical and experimental study of photon pathsApplied Optics, 1995
- Total Forearm Blood Flow as an Indicator of Skeletal Muscle Blood Flow: Effect of Subcutaneous Adipose Tissue Blood FlowClinical Science, 1994
- Clinical Methods: The History, Physical, and Laboratory Examinations1990