Occlusion of the Ascenda Catheter in a Patient with Pump Twiddler’s Sydrome: A Case Report
Open Access
- 28 April 2018
- journal article
- Published by Brieflands in Anesthesiology and Pain Medicine
- Vol. 8 (2) , e65312
Abstract
Background: Intrathecal baclofen (ITB) therapy is an effective way to manage spasticity in numerous conditions, including multiple sclerosis, stroke, and cerebral palsy. While pump failure is a common complication of ITB, improvements in device design have led to reduction of complications. In particular, the Ascenda catheter from Medtronic, Inc. was designed to resist kinking and associated complications; indeed, no incidences of catheter twisting or occlusion have been reported in literature prior to this case. Case Report: We report a case of a 32-year old gentleman who presented to the clinic with symptoms of baclofen withdrawal 19 months after he had a programmable pump implanted for spasticity. During the diagnostic evaluation it was discovered that the patients pump had flipped in his abdominal pocket. He was taken to surgery to reorient the pump, during which time it was noted the catheter was tightly coiled on itself occluding flow. The twisted catheter was excised and replaced with a new segment.His symptoms subsequently resolved. Conclusions: Although catheter occlusions have subsided since the approval of the Ascenda catheter, pump twiddler’s syndrome remains a risk factor for this complication. This is the first report describing this syndrome in a patient with the Ascenda catheter. Keywords: Ascenda Catheter; Intrathecal Baclofen Therapy (ITB); Baclofen, Twiddler’s SyndromeKeywords
This publication has 14 references indexed in Scilit:
- Intrathecal Drug Delivery Systems (IDDS): The Implantable Systems Performance Registry (ISPR)Neuromodulation: Technology at the Neural Interface, 2016
- Comparison between an Ascenda and a silicone catheter in intrathecal baclofen therapy in pediatric patients: analysis of complicationsJournal of neurosurgery. Pediatrics, 2016
- Intrathecal Baclofen Pump Migration Into the Peritoneal Cavity: A Case ReportAnesthesiology and Pain Medicine, 2016
- Current perspectives on intrathecal drug deliveryJournal of Pain Research, 2014
- Twisted Catheter Causing Baclofen Pump Malfunction: A Case ReportCanadian Journal of Neurological Sciences / Journal Canadien des Sciences Neurologiques, 2012
- Review: Procedure- and Device-Related Complications of Intrathecal Baclofen Administration for Management of Adult Muscle Hypertonia: A ReviewNeurorehabilitation and Neural Repair, 2010
- Catheter Migration After Implantationan Intrathecal Baclofen Infusion Pump for Severe Spasticity: A Case ReportThe Kaohsiung Journal of Medical Sciences, 2008
- Device-related complications of long-term intrathecal drug therapy via implanted pumpsSpinal Cord, 2008
- The phenomenon of spasticity: a pathophysiological and clinical introduction to neuromodulation therapies2006
- The control of muscle tone, reflexes, and movementNeurology, 1980