Abstract
Statistical image reconstruction (SIR) methods have shown potential to substantially improve the image quality of low‐dose x‐ray computed tomography (CT) as compared to the conventional filtered back‐projection (FBP) method. According to the maximuma posteriori(MAP) estimation, theSIRmethods are typically formulated by an objective function consisting of two terms: (a) a data‐fidelity term that models imaging geometry and physical detection processes in projection data acquisition, and (b) a regularization term that reflects prior knowledge or expectations of the characteristics of the to‐be‐reconstructed image.SIRdesires accurate system modeling of data acquisition, while the regularization term also has a strong influence on the quality of reconstructed images. A variety of regularization strategies have been proposed forSIRin the past decades, based on different assumptions, models, and prior knowledge. In this paper, we review the conceptual and mathematical bases of these regularization strategies and briefly illustrate their efficacies inSIRof low‐doseCT.
Funding Information
  • National Natural Science Foundation of China (U1708261, 61571214, 81701690, 61701217)
  • Natural Science Foundation of Guangdong Province (2015A030313271)
  • National Institutes of Health (R01 EB020366)
  • American Cancer Society (RSG-13-326-01-CCE)

This publication has 188 references indexed in Scilit: