Abstract
Objectives: The In Vitro Diagnostics Regulation (IVDR) will be effective in May 2022 by which in-house developed tests need to apply to the general safety and performance requirements defined in Annex I of the IVDR ruling. Yet, article 16 from Annex I about software can be hard to interpret and implement, particularly as laboratories are unfamiliar with quality standards for software development. Methods: In this paper we provide recommendations on organizational structure, standards to use, and documentation, for IVDR compliant in-house software development. Results: A practical insight is offered into novel standard operating procedures using three examples: an Excel file with a formula to calculate the pharmacokinetics of tacrolimus and to calculate the new dose, a rule for automated diagnosis of acute kidney injury and a bioinformatics pipeline for DNA variant calling. Conclusions: We recommend multidisciplinary development teams supported by higher management, use of ISO-15189 in synergy with IEC-62304, and concise documentation that includes intended purpose, classification, requirement management, risk management, verification and validation, configuration management and references to clinical or performance evidence.