Evaluation of non-invasive sensors for monitoring core temperature
Authors:
- 22 March 2025
- journal article
- research article
- Published by Springer Nature in Journal of Clinical Monitoring and Computing
- Vol. 39 (6) , 1137-1142
Abstract
We evaluated the accuracy and precision of zero-heat flux (ZHF) and dual sensor (DS) non-invasive temperature probes in intensive care unit (ICU) patients undergoing hypothermic temperature control, hypothesizing that both devices would accurately estimate core temperature. In a single-center prospective cohort study, we enrolled 35 ICU patients and applied continuous, non-invasive ZHF and/or DS probes to the lateral forehead or anterior chest to collect 358 observations. Conditions potentially influencing temperature estimation were recorded. Using Bland–Altman analysis with multiple paired observations per individual, we compared the bias between non-invasive probes and direct core temperature measurements. Lin’s concordance coefficient (LCC) was computed to quantify precision. The mean bias between the ZHF probe and invasive temperature was + 0.98 °C; for the DS probe, it was − 2.19 °C. In hypothermic patients, the ZHF probe’s accuracy improved (bias + 0.28 °C, LCC 0.86), while the DS probe remained inaccurate (bias − 2.52 °C, LCC 0.07). Clinical confounders like vasoactive agents or temperature control devices did not consistently affect bias, accuracy, or precision. Neither the ZHF nor DS non-invasive probes provided sufficient accuracy or precision to guide clinical decisions in the ICU. These results contrast with previous studies reporting biases within ± 0.5 °C. However, the ZHF probe showed promising limited deviation, especially in hypothermic patients.Keywords
Funding Information
- NASA (NNX16AO69A)
This publication has 19 references indexed in Scilit:
- Temperature control after cardiac arrestResuscitation, 2023
- Comparison of zero heat flux and double sensor thermometers during spinal anaesthesia: a prospective observational studyJournal of Clinical Monitoring and Computing, 2022
- Accuracy and precision of zero-heat-flux temperature measurements with the 3M™ Bair Hugger™ Temperature Monitoring System: a systematic review and meta-analysisJournal of Clinical Monitoring and Computing, 2020
- Accuracy of the non-invasive Tcore™ temperature monitoring system to measure body core temperature in abdominal surgeryJournal of Clinical Monitoring and Computing, 2019
- Zero-heat-flux core temperature monitoring system: an observational secondary analysis to evaluate agreement with naso-/oropharyngeal probe during anesthesiaJournal of Clinical Monitoring and Computing, 2019
- The focus of temperature monitoring with zero-heat-flux technology (3M Bair-Hugger): a clinical study with patients undergoing craniotomyJournal of Clinical Monitoring and Computing, 2018
- Evaluation of the novel non-invasive zero-heat-flux Tcore™ thermometer in cardiac surgical patientsJournal of Clinical Monitoring and Computing, 2018
- The prehospital management of hypothermia — An up-to-date overviewInjury, 2018
- “Deep-forehead” temperature correlates well with blood temperatureCanadian Journal of Anesthesia/Journal canadien d'anesthésie, 2000
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986