Analytical Method of Estimating Chemometric Prediction Error
Authors:
- 1 May 1997
- journal article
- research article
- Published by SAGE Publications in Applied Spectroscopy
- Vol. 51 (5) , 725-732
Abstract
We present an analytical formula that estimates the uncertainty in concentrations predicted by linear multivariate calibration, particularly partial least-squares (PLS). We emphasize the analysis of spectroscopic data. The derivation addresses the important limit in which calibration error is negligible in comparison to noise in the prediction spectra. The formula is expressed in terms of standard PLS calibration parameters and the amplitude of spectral noise; it is therefore straightforward to evaluate. To test the formula, we performed PLS analysis upon simulated spectra and upon experimental Raman spectra of dissolved biological analytes in water. In each instance, the root-mean-squared error of prediction was compared to the estimate from the formula. Accurate uncertainty estimates were obtained in cases where calibration noise was lower than prediction noise, and surprisingly good estimates were obtained even when the noise levels were equal. By comparing measured and estimated uncertainties, we assessed the robustness of each PLS calibration model. The scaling of prediction uncertainty with the spectral signal-to-noise ratio is also discussed.Keywords
This publication has 12 references indexed in Scilit:
- Comparison of methods to determine chromophore concentrations from fluorescence spectra of turbid samplesLasers in Surgery and Medicine, 1996
- Aqueous Dissolved Gas Measurements Using Near-Infrared Raman SpectroscopyApplied Spectroscopy, 1995
- Application of a multivariate technique to Raman spectra for quantification of body chemicalsIEEE Transactions on Biomedical Engineering, 1995
- Assessment of the Quality of Latent Variable Calibrations Based on Monte Carlo SimulationsAnalytical Chemistry, 1994
- Noninvasive Blood Glucose Assay by Near-Infrared Diffuse Reflectance Spectroscopy of the Human Inner LipApplied Spectroscopy, 1993
- Rapid Near-Infrared Raman Spectroscopy of Human Tissue with a Spectrograph and CCD DetectorApplied Spectroscopy, 1992
- Comparison of multivariate calibration methods for quantitative spectral analysisAnalytical Chemistry, 1990
- Tensorial calibration: I. First‐order calibrationJournal of Chemometrics, 1988
- Partial least-squares methods for spectral analyses. 1. Relation to other quantitative calibration methods and the extraction of qualitative informationAnalytical Chemistry, 1988
- Estimation of prediction error for multivariate calibrationJournal of Chemometrics, 1988