Image Quality Optimization in 60 kVp Head-Neck CTA: A Comparative Study of FBP, ClearView, and ClearInfinity Reconstruction Algorithms
- 27 October 2025
- journal article
- research article
- Published by Springer Nature in Current Medical Science
- Vol. 45 (6) , 1504-1512
Abstract
Objective To compare the impact of different reconstruction algorithms on the image quality of 60 kVp head and neck CT angiography (CTA) using subjective and objective metrics, with a focus on vessel edge sharpness. Methods This prospective study enrolled 45 patients who underwent ultra-low-voltage (60 kVp) head and neck CTA. Image datasets were reconstructed with filtered back-projection (FBP), ClearView (CV) and ClearInfinity (CI) algorithms at low (30%), medium (50%), and high (70%) strengths. Image quality was assessed subjectively and objectively via the Kruskal‒Wallis test for multiple comparisons. Objective parameters, including edge rise slope (ERS) and edge rise distance (ERD), were analyzed via the Friedman test of multiple comparisons statistics. Results Subjective assessments favored the CI50 reconstruction algorithm, demonstrating superior or satisfactory results compared to the other algorithms, with significantly better vessel delineation, edge definition and diagnostic confidence (all P < 0.05). Objective analysis revealed that the CV50 and CV70 algorithms significantly reduced ERS and/or elevated ERD (both P < 0.05). However, the CI50 algorithm maintained comparable vessel edge sharpness (P > 0.05) across all evaluated head and neck vascular segments when compared with the FBP algorithm. Conclusions The CI50 reconstruction algorithm optimizes image quality in 60 kVp head and neck CTA. It provides vessel edge sharpness comparable to FBP while offering superior vessel delineation, edge definition, and diagnostic confidence compared to FBP and CV algorithm. These findings suggest that CI50 has the potential to improve diagnostic accuracy in low-dose vascular imaging.Keywords
Funding Information
- National Key Research and Development Program of China (2024YFC2419300)
- National Natural Science Foundation of China (82471967)
- Hubei Provincial Key Research and Development Program (2024BCB008)
- Natural Science Foundation of Hubei Province (2025AFB733)
This publication has 27 references indexed in Scilit:
- Artificial intelligence and its potential integration with the clinical practice of diagnostic imaging medical physicists: a reviewPhysical and Engineering Sciences in Medicine, 2025
- Computed Tomography of the HeadClinical Neuroradiology, 2023
- Deep learning image reconstruction algorithm reduces image noise while alters radiomics features in dual-energy CT in comparison with conventional iterative reconstruction algorithms: a phantom studyEuropean Radiology, 2022
- The impact of deep learning reconstruction on image quality and coronary CT angiography-derived fractional flow reserve valuesEuropean Radiology, 2022
- Sinogram-based deep learning image reconstruction technique in abdominal CT: image quality considerationsEuropean Radiology, 2021
- Application of Low Tube Voltage, Low-concentration Contrast Agent Using a 320-row CT in Coronary CT Angiography: Evaluation of Image Quality, Radiation Dose and Iodine IntakeCurrent Medical Science, 2020
- Image Quality of CT Angiography of Supra-Aortic ArteriesClinical Neuroradiology, 2018
- Image quality and radiation dose of brain computed tomography in children: effects of decreasing tube voltage from 120 kVp to 80 kVpPediatric Radiology, 2017
- Diagnostic performance of CT angiography in neck vessel trauma: systematic review and meta-analysisEmergency Radiology, 2016
- Computed tomography angiography of the carotid arteries at low kV settings: a prospective randomised trial assessing radiation dose and diagnostic confidenceEuropean Radiology, 2011