Characterisation of tissue shrinkage during microwave thermal ablation
Authors:
- Laura FarinaLaura Farina
- Noam WeissNWNoam Weiss
- Yitzhak NissenbaumYNYitzhak Nissenbaum
- Marta CavagnaroMCMarta Cavagnaro
- Vanni LoprestoVLVanni Lopresto
- Rosanna PintoRPRosanna Pinto
- Nevio TosorattiNTNevio Tosoratti
- Claudio AmabileCAClaudio Amabile
- Simone CassarinoSCSimone Cassarino
- S. Nahum GoldbergSGS. Nahum Goldberg
- 17 October 2014
- journal article
- Published by Taylor & Francis in International Journal of Hyperthermia
- Vol. 30 (7) , 419-428
Abstract
The aim of this study was to characterise changes in tissue volume during image-guided microwave ablation in order to arrive at a more precise determination of the true ablation zone. The effect of power (20-80 W) and time (1-10 min) on microwave-induced tissue contraction was experimentally evaluated in various-sized cubes of ex vivo liver (10-40 mm ± 2 mm) and muscle (20 and 40 mm ± 2 mm) embedded in agar phantoms (N = 119). Post-ablation linear and volumetric dimensions of the tissue cubes were measured and compared with pre-ablation dimensions. Subsequently, the process of tissue contraction was investigated dynamically during the ablation procedure through real-time X-ray CT scanning. Overall, substantial shrinkage of 52-74% of initial tissue volume was noted. The shrinkage was non-uniform over time and space, with observed asymmetry favouring the radial (23-43 % range) over the longitudinal (21-29%) direction. Algorithmic relationships for the shrinkage as a function of time were demonstrated. Furthermore, the smallest cubes showed more substantial and faster contraction (28-40% after 1 min), with more considerable volumetric shrinkage (>10%) in muscle than in liver tissue. Additionally, CT imaging demonstrated initial expansion of the tissue volume, lasting in some cases up to 3 min during the microwave ablation procedure, prior to the contraction phenomenon. In addition to an asymmetric substantial shrinkage of the ablated tissue volume, an initial expansion phenomenon occurs during MW ablation. Thus, complex modifications of the tissue close to a radiating antenna will likely need to be taken into account for future methods of real-time ablation monitoring.Keywords
This publication has 25 references indexed in Scilit:
- Experimental characterisation of the thermal lesion induced by microwave ablationInternational Journal of Hyperthermia, 2014
- Computational modelling of microwave tumour ablationsInternational Journal of Hyperthermia, 2013
- Comparison of microwave ablation and multipolar radiofrequency ablation, both using a pair of internally cooled interstitial applicators: Results inex vivoporcine liversInternational Journal of Hyperthermia, 2011
- A Minimally Invasive Antenna for Microwave Ablation Therapies: Design, Performances, and Experimental AssessmentIEEE Transactions on Biomedical Engineering, 2011
- Principles of and Advances in Percutaneous AblationRadiology, 2011
- Pulmonary Thermal Ablation: Comparison of Radiofrequency and Microwave Devices by Using Gross Pathologic and CT Findings in a Swine ModelRadiology, 2009
- Microwave Ablation: Principles and ApplicationsRadioGraphics, 2005
- Creation of Radiofrequency Lesions in a Porcine ModelAmerican Journal of Roentgenology, 2000
- CT appearance of hepatic tumors after microwave coagulation therapy.American Journal of Roentgenology, 1998
- Ablation of liver tumors using percutaneous RF therapy.American Journal of Roentgenology, 1998