Comparison between diffusion-weighted MRI (DW-MRI) at 1.5 and 3 tesla: A phantom study
- 31 October 2013
- journal article
- research article
- Published by Wiley in Journal of Magnetic Resonance Imaging
- Vol. 40 (3) , 682-690
Abstract
Purpose To compare DW‐MRI between 1.5 and 3 Tesla (T) in terms of image quality, apparent diffusion coefficient (ADC), reproducibility, lesion‐to‐background contrast and signal‐to‐noise ratio (SNR), using a test object. Materials and Methods A spherical diffusion phantom was used for qualitatively assessing image quality and performing quantitative measurements between the two field strengths. Results Distortions and signal losses degraded image quality at 3T even when the protocols were optimized for minimum TE. The ADC, in the majority of the phantom compartments, was significantly different between 1.5T and 3T (P < 0.009), while the average coefficient of variation, excluding the phantom compartments affected by artifacts, was b = 1000 s/mm2 and comparable contrast was achieved at 3T with higher b‐values. The SNR gain at 3T could, in theory, be balanced by the increased number of signal excitations one can accommodate at 1.5T to perform DW‐MRI within the same acquisition time and possibly improved image quality, when 3T systems with no parallel transmission are used. Conclusion Further phantom and in vivo studies are required to investigate the utility of DW‐MRI at 3T, if image quality and acquisition times comparable to the ones from 1.5T are assumed. J. Magn. Reson. Imaging 2014;40:682–690.Keywords
This publication has 36 references indexed in Scilit:
- Diffusion-Weighted Magnetic Resonance Imaging as a Cancer Biomarker: Consensus and RecommendationsNeoplasia, 2009
- Technology Insight: water diffusion MRI—a potential new biomarker of response to cancer therapyNature Clinical Practice Oncology, 2008
- Clinical advantages of 3.0T MRI over 1.5TEuropean Journal of Radiology, 2008
- Artifacts in 3-T MRI: Physical background and reduction strategiesEuropean Journal of Radiology, 2008
- Diffusion Magnetic Resonance Imaging: A Biomarker for Treatment Response in OncologyJournal of Clinical Oncology, 2007
- Diffusion-Weighted MRI in the Body: Applications and Challenges in OncologyAmerican Journal of Roentgenology, 2007
- Imaging artifacts at 3.0TJournal of Magnetic Resonance Imaging, 2006
- Diffusion-weighted and Perfusion MR Imaging for Brain Tumor Characterization and Assessment of Treatment ResponseRadiology, 2006
- Functional diffusion map: A noninvasive MRI biomarker for early stratification of clinical brain tumor responseProceedings of the National Academy of Sciences, 2005
- Fast magnetic resonance diffusion‐weighted imaging of acute human strokeNeurology, 1992