Cellular Electrophysiological Changes Induced In Vitro by Radiofrequency Current: Comparison with Electrical Ablation
Authors:
- 1 February 1995
- journal article
- research article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 18 (2) , 323-333
Abstract
The purpose of this study was to examine the cellular electrophysiological effects of radiofrequency energy delivery in an in vitro canine epicardial preparation and compare the effects of those of high energy electrical ablation in a similar preparation. Ten joules of direct current energy or 40 volts of radiofrequency energy were delivered by a 6 French 2-mm tip catheter to the epicardial surface of 2 x 3 cm epicardial strips superfused with Tyrode's solution. Direct current energy delivery produced a crater and central zone of necrosis surrounded by a border zone of viable but damaged tissue that extended up to 10-12 mm from the site of energy delivery. Cellular electrophysiological abnormalities that included a less negative resting membrane potential, decreased peak dV/dT, decreased action potential amplitude, and decreased action potential duration (APD) were approximately linearly related to the distance from the crater edge. In addition, viable and inexcitable cells were frequently interspersed. Between 2 and 5 mm from the crater edge, 36.4% of the cells were inexcitable whereas others displayed normal action potential characteristics. In contrast, radiofrequency current produced a central zone of necrosis surrounded by a smaller border zone. Cellular damage that was qualitatively similar to that produced by direct current energy extended only up to 6-8 mm from the edge of the crater. In addition, severe abnormalities were noted in intracellular potentials recorded within 2 mm of the ablation site, and only minor abnormalities further away. Lesions were relatively homogeneous. Between 2 and 5 mm from the ablation site only 2.6% of the cells were inexcitable (P < 0.05 vs direct current). In conclusion, radiofrequency current produces lesions that are smaller and more homogeneous than those produced by direct current ablation. Although the border zone is small, a region of partially depolarized but viable myocardium is present after radiofrequency current energy delivery. These findings provide a cellular basis for several clinical observations that have been made following radiofrequency current energy delivery.Keywords
This publication has 17 references indexed in Scilit:
- Predictors of in-hospital mortality after DC catheter ablation of atrioventricular junction. Results of a prospective, international, multicenter study.Circulation, 1991
- Advances in Applications of Radiofrequency Current to Catheter Ablation TherapyPacing and Clinical Electrophysiology, 1991
- An Automatic Microcomputer System for Analysis of Monophasic Action PotentialsPacing and Clinical Electrophysiology, 1990
- Tissue Heating During Radiofrequency Catheter Ablation: A Thermodynamic Model and Observations in Isolated Perfused and Superfused Canine Right Ventricular Free WallPacing and Clinical Electrophysiology, 1989
- Catheter ablation of accessory pathways with a direct approach. Results in 35 patients.Circulation, 1988
- The cellular electrophysiologic changes induced by ablation: comparison between argon laser photoablation and high-energy electrical ablation.Circulation, 1987
- Catheter ablation of ventricular tachycardia with intracardiac shocks: results in 33 patients.Circulation, 1987
- The cellular electrophysiologic changes induced by high-energy electrical ablation in canine myocardium.Circulation, 1986
- Catheter ablation of the atrioventricular junction: a report of the percutaneous mapping and ablation registry.Circulation, 1984
- Myocardial injury and induction of arrhythmia by direct current shock delivered via endocardial catheters in dogs.Circulation, 1984