Fiberoptic Intraparenchymal Brain Pressure Monitoring with the Camino V420 Monitor: Reflections on Our Experience in 163 Severely Head-Injured Patients
- 1 April 2002
- journal article
- research article
- Published by SAGE Publications in Journal of Neurotrauma
- Vol. 19 (4) , 439-448
Abstract
To assess the safety and accuracy of the Camino intraparenchymal sensor, we prospectively evaluated hemorrhagic complications, zero-drift, infection, and system malfunction in 163 patients monitored after a severe head injury. Mean duration of intracranial pressure (ICP) monitoring was 5 ± 2.2 days (range: 12 h to 11 days). Of the 141 patients with a control CT scan, four showed a 1-2-cc collection of blood at the catheter's end. When removed, the sensors underread the true ICP value (negative zero-drift) in 80 of the 126 sensors evaluated (63.5%). Fourteen sensors showed no zero-drift, and 32 sensors overread the true ICP value (positive zero-drift) (median: -1 mm Hg; interquartile range: -4 to +1 mm Hg). No significant relationship was found between zero-drift, the surgeon who implanted the sensor, intracranial hypertension, or duration of ICP monitoring. No clinical infections could be attributed to the devices. Sixteen patients (9.8%) required more than one ICP sensor due to malfunctioning of the system. In conclusion, continuous ICP monitoring using the Camino intraparenchymal sensor has a low complication rate. However, this sensor may underread the real ICP values in a high number of patients. The lack of correlation between duration of ICP monitoring and zero-drift suggests that, contrary to the recommendations of other reports, the intraparenchymatous Camino sensor can provide reliable readings after the fifth day of use.Keywords
This publication has 19 references indexed in Scilit:
- Prospective study of zero drift in fiberoptic pressure monitors used in clinical practiceJournal of Neurosurgery, 1997
- EditorialJournal of Neurotrauma, 1996
- Systematic overestimation of intracranial pressure measured using a Camino pressure monitor.Journal of Neurology, Neurosurgery & Psychiatry, 1996
- Laboratory Testing of Three Intracranical Pressure Microtransducers: Technical ReportNeurosurgery, 1996
- Survey of critical care management of comatose, head-injured patients in the United StatesCritical Care Medicine, 1995
- An Evaluation of the Camino Ventricular Bolt System in Clinical PracticeNeurosurgery, 1993
- Complications of intracranial pressure monitoring in fulminant hepatic failureThe Lancet, 1993
- Monitoring of Brain Tissue Pressure with a Fiberoptic DeviceNeurosurgery, 1992
- Monitorage de la pression intracrânienne par capteur intraparenchymateux à fibres optiques : aspects techniques et fiabilité cliniqueAnnales Françaises d'Anesthésie et de Réanimation, 1992
- Evaluation of a fiberoptic intracranial pressure monitorJournal of Neurosurgery, 1990