The Effect of Feedback During Training Sessions on Learning Pattern-Recognition-Based Prosthesis Control
Open Access
- 20 August 2019
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Neural Systems and Rehabilitation Engineering
- Vol. 27 (10) , 2087-2096
Abstract
Human-machine interfaces have not yet advanced to enable intuitive control of multiple degrees of freedom as offered by modern myoelectric prosthetic hands. Pattern Recognition (PR) control has been proposed to make human-machine interfaces in myoelectric prosthetic hands more intuitive, but it requires the user to generate high-quality, i.e., consistent and separable, electromyogram (EMG) patterns. To generate such patterns, user training is required and has shown promising results. However, how different levels of feedback affect effectivity in training differently, has not been established yet. Furthermore, a correlation between qualities of the EMG patterns (the focus of training) and user performance has not been shown yet. In this study, 37 able-bodied participants (mean age 21 years, 19 males) were recruited and trained PR control over five days. Three levels of feedback were tested for their effectiveness: no external feedback, visual feedback and visual feedback with coaching. Training resulted in improved performance from pre-to post-test with no interaction effect of feedback. Feedback did however affect the quality of the EMG patterns where people who did not receive external feedback generated higher amplitude patterns. A weak correlation was found between a principal component, composed of EMG amplitude and pattern variability, and performance. Our results show that training is highly effective in improving PR control regardless of feedback and that none of the quality metrics correlate with performance. We discuss how different levels of feedback can be leveraged to improve PR control training.Keywords
Funding Information
- European Unions’s Horizon 2020 Research and Innovation Programme (687795)
This publication has 36 references indexed in Scilit:
- User Training for Pattern Recognition-Based Myoelectric Prostheses: Improving Phantom Limb Movement Consistency and DistinguishabilityIEEE Transactions on Neural Systems and Rehabilitation Engineering, 2013
- BioPatRec: A modular research platform for the control of artificial limbs based on pattern recognition algorithmsSource Code for Biology and Medicine, 2013
- A Training Strategy for Learning Pattern Recognition Control for Myoelectric ProsthesesJPO Journal of Prosthetics and Orthotics, 2013
- Patient Training for Functional Use of Pattern Recognition–Controlled ProsthesesJPO Journal of Prosthetics and Orthotics, 2012
- Instructions to Adopt an External Focus Enhance Muscular EnduranceResearch Quarterly for Exercise and Sport, 2011
- How changing the focus of attention affects performance, kinematics, and electromyography in dart throwingHuman Movement Science, 2010
- Effortless Motor Learning?: An External Focus of Attention Enhances Movement Effectiveness and Efficiency2010
- Targeted Muscle Reinnervation for Real-time Myoelectric Control of Multifunction Artificial ArmsJAMA, 2009
- Upper-Limb ProstheticsAmerican Journal of Physical Medicine & Rehabilitation, 2007
- EMG Activity as a Function of the Performer's Focus of AttentionJournal of Motor Behavior, 2004