The influence of luminance on the multifocal ERG
Authors:
- 16 November 2006
- journal article
- Published by Springer Nature in Documenta Ophthalmologica
- Vol. 113 (3) , 187-192
Abstract
To assess the efficacy of high luminance in increasing the amplitude of the multifocal electroretinogram (mfERG). We examined 5 male and 5 female volunteers in the age of 22–52 years (median 28 years). Three different stimulus luminance levels were applied: the bright areas of the stimulus pattern were set to 150, 300 and 500 cd/m2. We recorded the potentials via DTL electrodes using the VERIS Science 4.4 system with 61 hexagons, pupils were dilated. Analysis was based on the 5 ring averages. Across all hexagons and subjects, the response density (∼amplitude) rose by 20% when increasing the luminance by a factor of 3.3. The peak times decreased slightly with higher luminance, by less then 1.5 ms. Combining the present results with those from two previous studies, the gain (= relative amplitude increase for relative luminance increase) is close to 0.4 over a range of 56–700 cd/m2. The stimulus luminance range suggested in the mfERG guidelines seems well chosen.Keywords
This publication has 9 references indexed in Scilit:
- Luminance Dependence of Neural Components that Underlies the Primate Photopic ElectroretinogramInvestigative Ophthalmology & Visual Science, 2004
- The Multifocal ElectroretinogramJournal of Neuro-Ophthalmology, 2003
- Guidelines for basic multifocal electroretinography (mfERG)Documenta Ophthalmologica, 2003
- The effect of contrast and luminance on mfERG responses in a monkey model of glaucoma.2002
- Retinal origins of the primate multifocal ERG: implications for the human response.2002
- Multifocal electroretinogram: age-related changes for different luminance levelsGraefe's Archive for Clinical and Experimental Ophthalmology, 2002
- A Basic Investigation of Multifocal Electroretinogram Reproducibility and Effect of LuminanceJapanese Journal of Ophthalmology, 2000
- The field topography of ERG components in man—I. The photopic luminance responseVision Research, 1992
- Improved electrode for electroretinography.1979