Accuracy validation of a wearable IMU-based gait analysis in healthy female
Open Access
- 2 January 2024
- journal article
- research article
- Published by Springer Nature in BMC Sports Science, Medicine and Rehabilitation
- Vol. 16 (1) , 1-8
Abstract
Objective: The aim of this study was to assess the accuracy and test-retest reliability of a wearable inertial measurement unit (IMU) system for gait analysis in healthy female compared to a gold-standard optoelectronic motion capture (OMC) system. Methods: In our study, we collected data from 5 healthy young females. Participants were attached with markers from both the OMC system and the IMU system simultaneously. Data was collected when participants walked on a 7 m walking path. Each participant performed 50 repetitions of walking on the path. To ensure the collection of complete gait cycle data, a gait cycle was considered valid only if the participant passed through the center of the walking path at the same time that the OMC system detected a valid marker signal. As a result, 5 gait cycles that met the standards of the OMC system were included in the final analysis. The stride length, cadence, velocity, stance phase and swing phase of the spatio-temporal parameters were included in the analysis. A generalized linear mixture model was used to assess the repeatability of the two systems. The Wilcoxon rank-sum test for continuous variables was used to compare the mean differences between the two systems. For evaluating the reliability of the IMU system, we calculated the Intra-class Correlation Coefficient (ICC). Additionally, Bland-Altman plots were used to compare the levels of agreement between the two systems. Results: The measurements of Spatio-temporal parameters, including the stance phase (P = 0.78, 0.13, L-R), swing phase (P = 0.78, 0.13, L-R), velocity (P = 0.14, 0.13, L-R), cadence (P = 0.53, 0.22, L-R), stride length (P = 0.05, 0.19, L-R), by the IMU system and OMC system were similar. Which suggested that IMU and OMC systems could be used interchangeably for gait measurements. The intra-rater reliability showed an excellent correlation for the stance phase, swing phase, velocity and cadence (Intraclass Correlation Coefficient, ICC > 0.9) for both systems. However, the correlation of stride length was poor (ICC = 0.36, P = 0.34, L) to medium (ICC = 0.56, P = 0.22, R). Additionally, the measurements of IMU systems were repeatable. Conclusions: The results of IMU system and OMC system shown good repeatability. Wearable IMU system could analyze gait data accurately. In particular, the measurement of stance phase, swing phase, velocity and cadence showed excellent reliability. IMU system provided an alternative measurement to OMC for gait analysis. However, the measurement of stride length by IMU needs further consideration.Keywords
This publication has 35 references indexed in Scilit:
- Validity and repeatability of inertial measurement units for measuring gait parametersGait & Posture, 2017
- Examination of the reliability of an inertial sensor-based gait analysis systemBiomedical Engineering / Biomedizinische Technik, 2017
- Gait Partitioning Methods: A Systematic ReviewSensors, 2016
- Differences in gaze anticipation for locomotion with and without visionFrontiers in Human Neuroscience, 2015
- Quantified self and human movement: A review on the clinical impact of wearable sensing and feedback for gait analysis and interventionGait & Posture, 2014
- IMU-Based Joint Angle Measurement for Gait AnalysisSensors, 2014
- Gait Analysis Using Wearable SensorsSensors, 2012
- Reliability, repeatability and reproducibility: analysis of measurement errors in continuous variablesUltrasound in Obstetrics & Gynecology, 2008
- Assessment and validation of a simple automated method for the detection of gait events and intervalsGait & Posture, 2004
- Applying the right statistics: analyses of measurement studiesUltrasound in Obstetrics & Gynecology, 2003