A comparison of arterial lines and insertion techniques in critically ill patients
Authors:
- L. DoedensLDL. Doedens
- A. JacksonAJA. Jackson
- J. LipmanJ. Lipman
- 1 November 1994
- journal article
- clinical trial
- Published by Wiley in Anaesthesia
- Vol. 49 (11) , 968-973
Abstract
Summary: We compared three arterial line insertion techniques and two types of arterial catheters in 69 critically ill patients. Use of the direct‐puncture technique (method A) was associated with a significantly higher failure rate (23%) than use of a catheter with a separate guide wire (method B,‘classical’Seldinger technique, p < 0.001) or a catheter with an integral guide wire (method C,‘modified’Seldinger technique, p < 0.02). Operators randomly allocated to using method A took significantly longer to perform the procedure than those using method C (p < 0.01), used significantly more catheters (p < 0.0001) and made significantly more punctures in achieving a successful insertion than those using either methods B (p < 0.001) or C (p < 0.001). Both catheter types B and C (polyurethane) were significantly less likely to block, thus requiring re‐insertion, than catheter type A (Teflon) (p < 0.02, p < 00 01 respectively). We recommend the use of a‘classical’Seldinger technique (method B) for arterial line insertion in critically ill patients and the use of a polyurethane catheter, in preference to Teflon, to maximise catheter life after insertion.Keywords
This publication has 9 references indexed in Scilit:
- Direct or Modified Seldinger Guide Wire-Directed Technique for Arterial Catheter InsertionAnesthesia & Analgesia, 1993
- Arterial CatheterizationCritical Care Clinics, 1992
- Evaluation of Central Venous Catheter ThrombogenicityActa Anaesthesiologica Scandinavica, 1985
- Material thrombogenicity in central venous catheterization: a comparison between soft, antebrachial catheters of silicone elastomer and polyurethaneJournal of Parenteral and Enteral Nutrition, 1984
- On the Safety of Radial Artery CannulationAnesthesiology, 1983
- RADIAL ARTERY CANNULATIONBritish Journal of Anaesthesia, 1980
- Radial Arterial Function following Percutaneous Cannulation with 18− and 20-gauge CathetersAnesthesiology, 1977
- Percutaneous Radial-artery Cannulation—Increased Safety Using Teflon CathetersAnesthesiology, 1975
- Percutaneous Indwelling Radial-Artery Catheters for Monitoring Cardiovascular FunctionNew England Journal of Medicine, 1974