Comparison of the Batho, ETAR and Monte Carlo dose calculation methods in CT based patient models
Authors:
- F. C. P. du PlessisFPF. C. P. du Plessis
- C. A. WillemseCWC. A. Willemse
- M. G. LötterMLM. G. Lötter
- L. GoedhalsLGL. Goedhals
- 9 April 2001
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 28 (4) , 582-589
Abstract
This paper shows the contribution that Monte Carlo methods make in regard to dose distribution calculations in CT based patient models and the role it plays as a gold standard to evaluate other dose calculation algorithms. The EGS4 based BEAM code was used to construct a generic 8 MV accelerator to obtain a series of x‐ray field sources. These were used in the EGS4 based DOSXYZ code to generate beam data in a mathematical water phantom to set up a beam model in a commercial treatment planning system (TPS), CADPLAN V.2.7.9. Dose distributions were calculated with the Batho and ETAR inhomogeneity correction algorithms in head/sinus, lung, and prostate patient models for and open x‐ray beams. Corresponding dose distributions were calculated with DOSXYZ that were used as a benchmark. The dose comparisons are expressed in terms of 2D isodose distributions, percentage depth dose data, and dose difference volume histograms (DDVH's). Results indicated that the Batho and ETAR methods contained inaccuracies of 20%–70% in the maxillary sinus region in the head model. Large lung inhomogeneities irradiated with small fields gave rise to absorbed dose deviations of 10%–20%. It is shown for a field that DOSXYZ models lateral scatter in lung that is not present in the Batho and ETAR methods. The ETAR and Batho methods are accurate within 3% in a prostate model. We showed how the performance of these inhomogeneity correction methods can be understood in realistic patient models using validated Monte Carlo codes such as BEAM and DOSXYZ.Keywords
This publication has 27 references indexed in Scilit:
- Dose calculations using convolution and superposition principles: The orientation of dose spread kernels in divergent x-ray beamsMedical Physics, 1993
- Generation of photon energy deposition kernels using the EGS Monte Carlo codePhysics in Medicine & Biology, 1988
- Calculation and Application of Point Spread Functions for Treatment Planning with High Energy Photon BeamsActa Oncologica, 1987
- Differential pencil beam dose computation model for photonsMedical Physics, 1986
- Accuracy of lung dose calculations for large‐field irradiation with 6‐MV x raysMedical Physics, 1986
- Improved lung dose calculation using tissue-maximum ratios in the Batho correctionMedical Physics, 1984
- The Equivalent Tissue-Air Ratio Method for Making Absorbed Dose Calculations in a Heterogeneous MediumRadiology, 1978
- Corrections to absorbed dose calculations for tissue inhomogeneitiesMedical Physics, 1977
- First scatter to off-axis points and the Clarkson method (absorbed dose distributions)Physics in Medicine & Biology, 1973
- Scatter-air ratiosPhysics in Medicine & Biology, 1972