CT scanner x‐ray spectrum estimation from transmission measurements
- 28 January 2011
- journal article
- radiation imaging-physics
- Published by Wiley in Medical Physics
- Vol. 38 (2) , 993-997
Abstract
Purpose: In diagnostic CT imaging, multiple important applications depend on the knowledge of the x‐ray spectrum, including Monte Carlo dose calculations and dual‐energy material decomposition analysis. Due to the high photon flux involved, it is difficult to directly measure spectra from the x‐ray tube of a CT scanner. One potential method for indirect measurement involves estimating the spectrum from transmission measurements. The expectation maximization (EM) method is an accurate and robust method to solve this problem. In this article, this method was evaluated in a commercial CT scanner.Methods: Two step‐wedges (polycarbonate and aluminum) were used to produce different attenuation levels. Transmission measurements were performed on the scanner and the measured data from the scanner were exported to an external computer to calculate the spectra. The EM method was applied to solve the equations that represent the attenuation processes of polychromatic x‐ray photons. Estimated spectra were compared to the spectra simulated using a software provided by the manufacturer of the scanner. To test the accuracy of the spectra, a verification experiment was performed using a phantom containing different depths of water. The measured transmission data were compared to the transmission values calculated using the estimated spectra.Results: Spectra of 80, 100, 120, and 140 kVp from a dual‐source CT scanner were estimated. The estimated and simulated spectra were well matched. The differences of mean energies were less than 1 keV. In the verification experiment, the measured and calculated transmission values were in excellent agreement.Conclusions: Spectrum estimation using transmission data and the EM method is a quantitatively accurate and robust technique to estimate the spectrum of a CT system. This method could benefit studies relying on accurate knowledge of the x‐ray spectra from CT scanner.Keywords
Funding Information
- National Institutes of Health (EB007986)
- National Institutes of Health (DK083007)
This publication has 18 references indexed in Scilit:
- Calculation of 10 MV x-ray spectra emitted by a medical linear accelerator using the BFGS quasi-Newton methodPhysics in Medicine & Biology, 2006
- Computer algebra for x-ray spectral reconstruction between 6 and 25 MVMedical Physics, 2001
- X‐ray spectra estimation using attenuation measurements from 25 kVp to 18 MVMedical Physics, 1999
- Diagnostic x‐ray spectra: A comparison of spectra generated by different computational methods with a measured spectrumMedical Physics, 1998
- An accurate method for computer‐generating tungsten anode x‐ray spectra from 30 to 140 kVMedical Physics, 1997
- Estimation of a photon energy spectrum for a computed tomography scannerMedical Physics, 1997
- Flat and curved crystal spectrography for mammographic X-ray sourcesBritish Journal of Radiology, 1996
- Compton spectroscopy in the diagnostic X-ray energy range: I. Spectrometer designPhysics in Medicine & Biology, 1989
- A Compton scattering spectrometer for determining X-ray photon energy spectraPhysics in Medicine & Biology, 1987
- Computation of bremsstrahlung X-ray spectra and comparison with spectra measured with a Ge(Li) detectorPhysics in Medicine & Biology, 1979