Determination of hemolysis index thresholds for biochemical tests on Siemens Advia 2400 chemistry analyzer
Open Access
- 19 February 2019
- journal article
- research article
- Published by Wiley in Journal of Clinical Laboratory Analysis
- Vol. 33 (4) , e22856
Abstract
Background: In vitro hemolysis is still the most common source of pre‐analytical nonconformities. This study aimed to investigate the hemolytic effects on commonly used biochemical tests as well as to determine the hemolysis index (HI) thresholds on Siemens Advia 2400 chemistry analyzer.Methods: Peripheral blood samples were collected from forty healthy volunteers. Hemolysis was achieved using syringes. Five hemolysis levels were produced including the no hemolysis group, slight hemolysis group, mild hemolysis group, moderate hemolysis group, and heavy hemolysis group. We then used the bias from baseline (no hemolysis) and HI to construct regression functions. The HI corresponding to the bias limits was considered as HI thresholds. We chose the total allowable error (TAE) as the bias limit.Results: Of the twenty‐eight analytes, ten analytes had clinical significance. Creatine kinase‐MB, creatine kinase, potassium, aspartate aminotransferase, and hydroxybutyrate dehydrogenase were all positively affected; the corresponding HI threshold was 45.2, 99.96, 4.07, 10.16, and 7.94, respectively. Lactate dehydrogenase was also positively interfered, but we failed to calculate the HI threshold. Total bile acid, uric acid, and sodium were all negatively affected, and the HI threshold was 42.23, 500 and 501.8, respectively. Glucose was also negatively interfered, but it failed to achieve the HI threshold.Conclusions: When the HI value was higher than its threshold, the corresponding analyte was considered inappropriate for reporting. The implementation of the assay‐specific HI thresholds could provide an accurate method to identify analytes interfered by hemolysis, which would improve clinical interpretations and further boost laboratory quality by reducing errors associated with hemolysis.Keywords
Funding Information
- Tianjin Science and Technology Project (15ZXLCSY0040)
- the Basic research project of Logistics College of Armed Police Forces (WHJ2016024)
This publication has 22 references indexed in Scilit:
- Critical review and meta-analysis of spurious hemolysis in blood samples collected from intravenous cathetersBiochemia Medica, 2013
- Speed of sample transportation by a pneumatic tube system can influence the degree of hemolysisClinical Chemistry and Laboratory Medicine (CCLM), 2012
- Hemoglobin, bilirubin, and lipid interference on Roche Cobas 6000 assaysClinica Chimica Acta, 2012
- Influence of in vitro hemolysis on hematological testing on Advia 2120International Journal of Laboratory Hematology, 2011
- Evaluation of the interference of hemoglobin, bilirubin, and lipids on Roche Cobas 6000 assaysClinica Chimica Acta, 2011
- The prevalence of preanalytical errors in a Croatian ISO 15189 accredited laboratoryClinical Chemistry and Laboratory Medicine (CCLM), 2010
- Hemolysis detection and management of hemolysed specimensBiochemia Medica, 2010
- To err is human. To misdiagnose might be deadlyClinical Biochemistry, 2010
- Multicenter evaluation of the hemolysis index in automated clinical chemistry systemsClinical Chemistry and Laboratory Medicine (CCLM), 2009
- Haemolysis: an overview of the leading cause of unsuitable specimens in clinical laboratoriesClinical Chemistry and Laboratory Medicine (CCLM), 2008