Accuracy and precision of hemoglobin point‐of‐care testing during major pediatric surgery
Authors:
- 27 July 2011
- journal article
- research article
- Published by Wiley in International Journal of Laboratory Hematology
- Vol. 34 (1) , 86-90
Abstract
Summary: Introduction: The aim of the study was to compare accuracy and reproducibility of four point‐of‐care testing (POCT) devices (GEM® Premier 3000, ABL 800 flex, GEM®OPL™, HemoCue® B‐Hemoglobin) for hemoglobin (Hb) analyzes as compared with the reference laboratory method (Sysmex XE 2100) in children undergoing major surgery.Methods: Pediatric patients undergoing craniofacial, spine, hip, or cancer surgery were included. Blood samples for Hb testing were taken at several intraoperative time points and generally withdrawn from the arterial catheter, if accessible.Results: A total of 256 blood samples were taken intraoperatively from 71 pediatric patients. All POCT devices showed very small bias (maximum −0.46 g/dL) to reference method as well as very good reproducibility (maximum coefficient of variation of 0.99%). However, in two cases (HemoCue), potential clinical relevant differences were observed beyond a range of 2 g/dL.Conclusion: All POCT devices tested and operated by trained staff for hemoglobinometry showed reliable test results. They all allow for simple, fast, and precise bedside determination of hemoglobin concentration in the intraoperative setting.Keywords
This publication has 13 references indexed in Scilit:
- Estimation of hemoglobin in blood donors: A comparative study using hemocue and cell counterTransfusion and Apheresis Science, 2010
- Evaluation of four rapid methods for hemoglobin screening of whole blood donors in mobile collection settingsTransfusion and Apheresis Science, 2007
- Hemoglobin Test Result Variability and Cost Analysis of Eight Different Analyzers During Open Heart SurgeryThe Journal of ExtraCorporeal Technology, 2007
- Reduction in Red Blood Cell Transfusions Using a Bedside Analyzer in Extremely Low Birth Weight InfantsJournal of Perinatology, 2004
- Comparison between the HemoCue® and an automated counter for measuring hemoglobinRevista de Saúde Pública, 2004
- Evaluation of the blood gas analyzer GEM(r) PREMIER(tm) 3000Clinical Chemistry and Laboratory Medicine (CCLM), 2004
- Accuracy of point‐of‐care‐testing (POCT) for determining hemoglobin concentrationsActa Anaesthesiologica Scandinavica, 2002
- Blood Analysis at the Point of Care: Issues in Application for Use in Critically Ill PatientsAACN Advanced Critical Care, 2002
- Evaluation of the HemoCue ® for measuring intra‐operative haemoglobin concentrations: a comparison with the Coulter Max‐M ®Anaesthesia, 1998
- Limits of performance for measurement of blood haemoglobin concentration, number concentration of erythrocytes, and erythrocyte volume fraction in interlaboratory trials: Results of the Finnish proficiency testing programme in haematologyScandinavian Journal of Clinical and Laboratory Investigation, 1980