Closed-Loop Control and Advisory Mode Evaluation of an Artificial Pancreatic β Cell: Use of Proportional-Integral-Derivative Equivalent Model-Based Controllers
Open Access
- 1 July 2008
- journal article
- research article
- Published by SAGE Publications in Journal of Diabetes Science and Technology
- Vol. 2 (4) , 636-644
Abstract
Background: Using currently available technology, it is possible to apply modern control theory to produce a closed-loop artificial β cell. Novel use of established control techniques would improve glycemic control, thereby reducing the complications of diabetes. Two popular controller structures, proportional-integral-derivative (PID) and model predictive control (MPC), are compared first in a theoretical sense and then in two applications. Methods: The Bergman model is transformed for use in a PID equivalent model-based controller. The internal model control (IMC) structure, which makes explicit use of the model, is compared with the PID controller structure in the transfer function domain. An MPC controller is then developed as an optimization problem with restrictions on its tuning parameters and is shown to be equivalent to an IMC controller. The controllers are tuned for equivalent performance and evaluated in a simulation study as a closed-loop controller and in an advisory mode scenario on retrospective clinical data. Results: Theoretical development shows conditions under which PID and MPC controllers produce equivalent output via IMC. The simulation study showed that the single tuning parameter for the equivalent controllers relates directly to the closed-loop speed of response and robustness, an important result considering system uncertainty. The risk metric allowed easy identification of instances of inadequate control. Results of the advisory mode simulation showed that suitable tuning produces consistently appropriate delivery recommendations. Conclusion: The conditions under which PID and MPC are equivalent have been derived. The MPC framework is more suitable given the extensions necessary for a fully closed-loop artificial β cell, such as consideration of controller constraints. Formulation of the control problem in risk space is attractive, as it explicitly addresses the asymmetry of the problem; this is done easily with MPC.Keywords
This publication has 22 references indexed in Scilit:
- Utilization of a Computerized Intravenous Insulin Infusion Program to Control Blood Glucose in the Intensive Care UnitDiabetes Technology & Therapeutics, 2007
- A Critical Assessment of Algorithms and Challenges in the Development of a Closed-Loop Artificial PancreasDiabetes Technology & Therapeutics, 2005
- Modeling Insulin Action for Development of a Closed-Loop Artificial PancreasDiabetes Technology & Therapeutics, 2005
- Management of diabetes using adaptive controlInternational Journal of Adaptive Control and Signal Processing, 2004
- Intensive blood-glucose control with sulphonylureas or insulin compared with conventional treatment and risk of complications in patients with type 2 diabetes (UKPDS 33)The Lancet, 1998
- The Effect of Intensive Treatment of Diabetes on the Development and Progression of Long-Term Complications in Insulin-Dependent Diabetes MellitusNew England Journal of Medicine, 1993
- Internal model control: PID controller designIndustrial & Engineering Chemistry Process Design and Development, 1986
- Internal model control. A unifying review and some new resultsIndustrial & Engineering Chemistry Process Design and Development, 1982
- Physiologic evaluation of factors controlling glucose tolerance in man: measurement of insulin sensitivity and beta-cell glucose sensitivity from the response to intravenous glucose.Journal of Clinical Investigation, 1981
- Model predictive heuristic controlAutomatica, 1978