Performance evaluation of the new CoaguChek XS system compared with the established CoaguChek system by patients experienced in INR-self management
Authors:
- 1 January 2007
- journal article
- research article
- Published by Georg Thieme Verlag KG in Thrombosis and Haemostasis
- Vol. 97 (02) , 310-314
Abstract
The aim of the study was to analyze the concordance of INR values obtained by educated lay users with those obtained by professionals and to determine the imprecision of the new system. The new CoaguChek® XS system was tested in a user study over six weeks at four study centres in Austria, Denmark and Germany. Seventy-five patients receiving oral anticoagulant therapy were enrolled in the study. The INR results in capillary whole blood taken by professionals and by patients using the Coagu- Chek XS system were similar, and the mean relative bias was < 1%. The imprecision of the CoaguChek XS system calculated from duplicate testing is low (<6%) and slightly better than for the established CoaguChek S system. The INR results measured during the home testing phase correlated quite well between the established CoaguChek S system and the new CoaguChek XS system with a mean bias of 0.14 INR. This is a remarkably low bias taking into consideration that more than 30 different test strip lots were applied. A questionnaire was filled out by all patients to assess their personal impression. It revealed that patients were very satisfied with the new system and found it easy to operate. The results demonstrate that the agreement between professional and patient INR results for the new Coagu- Chek XS system was excellent and that INR values can be determined by lay users as well as by professionals. The instrument is very well accepted by the patients and their satisfaction even increased after four weeks practice at home.Keywords
Funding Information
- Roche Diagnostics GmbH
This publication has 13 references indexed in Scilit:
- Self-monitoring of oral anticoagulation: a systematic review and meta-analysisThe Lancet, 2006
- Short-term venous stasis influences routine coagulation testingBlood Coagulation & Fibrinolysis, 2005
- Poor Agreement among Prothrombin Time International Normalized Ratio Methods: Comparison of Seven Commercial ReagentsClinical Chemistry, 2005
- Importance of Device Evaluation for Point-of-Care Prothrombin Time International Normalized Ratio Testing ProgramsMayo Clinic Proceedings, 2005
- Agreement of a new whole-blood PT/INR test using capillary samples with plasma INR determinationsThrombosis Research, 2005
- Accuracy and clinical usefulness of the near‐patient testing CoaguChek S international normalised ratio monitor in rural medical practiceAustralian Journal of Rural Health, 2004
- Systematic review of studies of self-management of oral anticoagulationThrombosis and Haemostasis, 2004
- INR Self-Management Permits Lower Anticoagulation Levels After Mechanical Heart Valve ReplacementCirculation, 2003
- Multicenter evaluation of a new capillary blood prothrombin time monitoring systemBlood Coagulation & Fibrinolysis, 1995
- A New Biometrical Procedure for Testing the Equality of Measurements from Two Different Analytical Methods. Application of linear regression procedures for method comparison studies in Clinical Chemistry, Part IClinical Chemistry and Laboratory Medicine (CCLM), 1983