Clinical and electrocardiographic predictors of T wave oversensing in patients with subcutaneous ICD

Authors:

Abstract
Background: T wave oversensing (TWOS) is a major drawback of the subcutaneous implantable cardioverter defibrillator (S‐ICD). Data on predictors of TWOS in S‐ICD recipients are limited.We sought to investigate predictors of TWOS in a cohort of patients receiving an S‐ICD at our institution.Methods: S‐ICD recipients at our center were identified retrospectively and stratified based on the presence or absence of TWOS. Clinical and electrocardiographic parameters were collected and compared between the 2 groups.Results: Ninety‐two patients underwent an S‐ICD implantation at our institution between April 2010 and January 2015. Six (6.5%) patients had TWOS. These patients were younger (38.1±13.7 vs. 52.3±16.1 years, p=0.04) and had higher left ventricle ejection fractions (48.5±14.9% vs. 28.4±12.2%, pp=0.02).Conclusion: In this study, younger age, higher ejection fraction, and lower QRS amplitude were associated with TWOS. These findings could help identify patients referred for S‐ICD at high‐risk of TWOS.