Propagation of calibration errors in prospective motion correction using external tracking
Open Access
- 2 October 2013
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 72 (2) , 381-388
Abstract
Purpose: Prospective motion correction of MRI scans using an external tracking device (such as a camera) is becoming increasingly popular, especially for imaging of the head. In order for external tracking data to be transformed into the MR scanner reference frame, the pose (i.e., position and orientation) of the camera relative to the scanner—or cross‐calibration—must be accurate. In this study, we investigated how errors in cross‐calibration affect the accuracy of motion correction feedback in MRI.Theory and Methods: An operator equation is derived describing how calibration errors relate to errors in applied motion compensation. By taking advantage of spherical symmetry and performing a Taylor approximation for small rotation angles, a closed form expression and upper limit for the residual tracking error is provided.Results: Experiments confirmed theoretical predictions of a bilinear dependence of the residual rotational component on the calibration error and the motion performed, modulated by a sinusoidal dependence on the angle between the calibration error axis and motion axis. The residual translation error is bounded by the sum of the rotation angle multiplied by the translational calibration error plus the linear head displacement multiplied by the calibration error angle.Conclusion: The results make it possible to calculate the required cross‐calibration accuracy for external tracking devices for a range of motions. Scans with smaller expected movements require less accuracy in cross‐calibration than scans involving larger movements. Typical clinical applications require that the calibration accuracy is substantially below 1 mm and 1°. Magn Reson Med 72:381–388, 2014.Keywords
Funding Information
- National Institutes of Health (1 R01 DA021146, U54 NS 056833, K02-DA16991)
This publication has 10 references indexed in Scilit:
- Fast noniterative calibration of an external motion tracking deviceMagnetic Resonance in Medicine, 2013
- Measurement and Correction of Microscopic Head Motion during Magnetic Resonance Imaging of the BrainPLOS ONE, 2012
- An embedded optical tracking system for motion-corrected magnetic resonance imaging at 7TMagnetic Resonance Materials in Physics, Biology and Medicine, 2012
- Real‐time optical motion correction for diffusion tensor imagingMagnetic Resonance in Medicine, 2011
- Navigator accuracy requirements for prospective motion correctionMagnetic Resonance in Medicine, 2009
- Magnetic resonance imaging of freely moving objects: prospective real-time motion correction using an external optical motion tracking systemNeuroImage, 2006
- Hand-Eye Calibration Using Dual QuaternionsThe International Journal of Robotics Research, 1999
- Extrinsic calibration of a vision sensor mounted on a robotIEEE Transactions on Robotics and Automation, 1992
- Adaptive technique for high-definition MR imaging of moving structures.Radiology, 1989
- A new technique for fully autonomous and efficient 3D robotics hand/eye calibrationIEEE Transactions on Robotics and Automation, 1989