Ceramic capillary electrophoresis chip for the measurement of inorganic ions in water samples
- 5 February 2010
- journal article
- research article
- Published by Royal Society of Chemistry (RSC) in The Analyst
- Vol. 135 (5) , 965-970
Abstract
We present a microchip capillary electrophoresis (CE) device build-up in low temperature co-fired ceramics (LTCC) multilayer technology for the analysis of major inorganic ions in water samples in less than 80 s. Contactless conductivity measurement is employed as a robust alternative to direct-contact conductivity detection schemes. The measurement electrodes are placed in a planar way at the top side of the CE chip and are realized by screen printing. Laser-cutting of channel and double-T injector structures is used to minimize irregularities and wall defects, elevating plate numbers per meter up to values of 110,000. Lowest limit of detection is 6 μM. The cost efficient LTCC module is attractive particularly for portable instruments in environmental applications because of its chemical inertness, hermeticity and easy three-dimensional integration capabilities of fluidic, electrical and mechanical components.Keywords
This publication has 32 references indexed in Scilit:
- A Monolithic Continuous-Flow Microanalyzer with Amperometric Detection Based on the Green Tape TechnologyAnalytical Chemistry, 2009
- Microchip electrophoresis in low‐temperature co‐fired ceramics technology with contactless conductivity measurementElectrophoresis, 2009
- Recent significant developments in detection and method development for the determination of inorganic ions by CEElectrophoresis, 2009
- Monitoring of cell cultures with LTCC microelectrode arrayAnalytical and Bioanalytical Chemistry, 2009
- Miniaturized Total Analysis Systems: Integration of Electronics and Fluidics Using Low-Temperature Co-Fired CeramicsAnalytical Chemistry, 2007
- A multi-sensor biological monitoring module built up in LTCC-technologyMicroelectronic Engineering, 2007
- LTCC-based microchips for the electrochemical detection of phenolic compoundsSensors and Actuators B: Chemical, 2006
- LTCC Microfluidic SystemInternational Journal of Applied Ceramic Technology, 2006
- Overview of low temperature co-fired ceramics tape technology for meso-system technology (MsST)Sensors and Actuators A: Physical, 2001
- Ceramic microchips for capillary electrophoresis–electrochemistryAnalytical Communications, 1999