Abstract
Quantification of bradykinin peptides in limited amounts of rat muscle tissue dialysate has been performed using a packed capillary LC‐ESI‐TOF‐MS method. The micro dialysate samples (450 μL) with added internal standard were loaded onto a 1 mm×5 mm loading column packed with 5 μm Kromasil C18 particles by a carrier solution of 0.1% formic acid in ACN/water (5:95, v/v) at a flow rate of 250 μL/min for online preconcentration of the analytes. Back‐flushed elution onto a 150 mm× 0.5 mm Zorbax C18 column packed with 5 μm particles was conducted using a linear solvent ACN/H2O gradient containing 0.1% formic acid. (Tyr8)‐bradykinin was used as an internal standard and was added to the dialysis sample prior to injection. Baseline separation of bradykinin, arg‐bradykinin and (tyr8)‐bradykinin was achieved within 10 min. Positive ESI was performed in the m/z range of 200–1300. The method was validated in the range 0.2–1.0 ng/mL dialysate, yielding correlation coefficients of 0.995 and 0.990 for bradykinin and arg‐bradykinin, respectively. The within‐assay and between‐assay precisions were between 4.3–9.6% and 6.2–10.6%, respectively. Both arg‐bradykinin and bradykinin were detected in dialysate from rat muscle tissue, at concentrations of 0.1 and 0.4 ng/mL for bradykinin and arg‐bradykinin, respectively, confirming the presence of arg‐bradykinin in rat muscles.

This publication has 19 references indexed in Scilit: