Zero-Bias Operation and Beam Energy Dependence of Diamond Sensors
Authors:
- 11 March 2013
- journal article
- Published by Trans Tech Publications, Ltd. in Key Engineering Materials
- Vol. 543, 426-430
Abstract
A prototype sensor able to work at zero bias has been fabricated using high quality single crystal diamond and two Schottky junctions. The detector's current in the dark is 3.2±0.2×10-14 A, while the current measured under irradiation with 1.6 Gymin-1 of 6 and 15 MV x-ray photons is 7.5±0.2×10-10 A. The sensitivity is strongly dependent on the applied voltage whereas at zero bias we have measured 37±2 nCGy-1, independent of the beam quality. The sensitivity increases with a sub-linear trend up to 100V, as the sensor photocurrent dependence on voltage. Moreover, the device shows very fast response on the rise and fall transients, with the signal stability under irradiation better than 0.3%. The linearity with the dose rate is spread over three orders of magnitude from 1.3×10-3 Gymin-1 to 1.2 Gymin-1 as measured with Co-60 photons.Keywords
This publication has 6 references indexed in Scilit:
- Zero-bias operation of polycrystalline chemically vapour deposited diamond films for Intensity Modulated Radiation TherapyDiamond and Related Materials, 2011
- Single crystal CVD diamond detector for high resolution dose measurement for IMRT and novel radiation therapy needsDiamond and Related Materials, 2010
- Single crystal diamond detector for radiotherapyJournal of Physics D: Applied Physics, 2010
- Operating parameters of CVD diamond detectors for radiation dosimetryNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2010
- Synthetic single crystal diamond dosimeters for conformal radiation therapy applicationDiamond and Related Materials, 2010
- Performance of Monocrystalline Diamond Radiation Detectors Fabricated Using TiW, Cr/Au and a Novel Ohmic DLC/Pt/Au Electrical ContactIEEE Transactions on Nuclear Science, 2009