Temperature Corrections for Ion Mobility Spectrometry
Author:
- 1 December 2001
- journal article
- research article
- Published by SAGE Publications in Applied Spectroscopy
- Vol. 55 (12) , 1653-1659
Abstract
The displacements of 30 peaks of eight chemicals over a wide range of temperatures, from 30 to 250 °C, with 20 °C increments, were studied. Experimental results show that the separations remain fairly constant and ion mobility spectra can be corrected by shifting all drift times by Δ t d (Δ T/T)[ a + b(Δ T/T)] where T is the absolute temperature and a and b are constants which depend on pressure and the instrumental parameters. Also, by using the reactant ion peak as an internal standard, a factor of k = ( t – t R )/ t R , has been defined, where t is the drift time, t R is the reactant ion drift time, and k is called the adjusted drift time. This factor is independent of instrumental parameters and increases linearly with temperature as: k = α( T 1/2 − 11.88). It can be normalized to a standard temperature of 298 K to give k 298 = 5.3827 k T /( T 1/2 − 11.88). The k 298 values were determined for the 30 peaks over a temperature range of 30–250 °C. These values show less sensitivity to temperature than their corresponding reduced mobilities. By transforming the entire drift time to the reduced adjusted drift time, k 298 , the peak shifts were corrected for all spectra recorded at temperatures (30–250 °C) such that the spectra superimposed.Keywords
This publication has 7 references indexed in Scilit:
- Design and optimization of a corona discharge ionization source for ion mobility spectrometryReview of Scientific Instruments, 2000
- Development of a Chemical Identification Algorithm for Gas Chromatography/Ion Mobility SpectrometrySAE International Journal of Advances and Current Practices in Mobility, 1998
- Prediction of Reduced Ion Mobility Constants of Organic Compounds from Molecular StructureAnalytical Chemistry, 1996
- Ion Mobility Spectrometry1993
- Ion Mobility Spectrometry in Analytical ChemistryCritical Reviews in Analytical Chemistry, 1990
- Mobility behavior and composition of hydrated positive reactant ions in plasma chromatography with nitrogen carrier gasAnalytical Chemistry, 1978
- Plasma Chromatography --A New Dimension for Gas Chromatography and Mass SpectrometryJournal of Chromatographic Science, 1970