Accuracy of ten isoflurane vaporisers in current clinical use
Authors:
- 13 July 2011
- journal article
- Published by Wiley in Anaesthesia
- Vol. 66 (8) , 682-688
Abstract
Summary: There is a lack of independent data regarding the accuracy of isoflurane vaporisers in current clinical use. We studied the accuracy of 10 isoflurane vaporisers (six Penlon Sigma Delta, four Datex‐Ohmeda Isotec 5) taken from our operating theatres and anaesthetic rooms. Isoflurane output was measured at a range of fresh gas flow rates using both 100% oxygen and a 1:2 oxygen:nitrous oxide mixture as carrier gas. Within a clinical range of fresh gas flow rates (1–6 l.min −1 ) the worst relative output deviation of each vaporiser was at least ±20%, with under‐delivery of isoflurane being commoner than over‐delivery. Relative output deviation ranged from −62% to +21%, which could lead to a risk of awareness or overdose. Despite the existence of an in‐house calibration schedule, the vaporisers we found in current clinical use under‐ or occasionally over‐delivered isoflurane to a dangerous and unpredictable extent.Keywords
This publication has 29 references indexed in Scilit:
- Evaluation of isoflurane and sevoflurane vaporizers over a wide range of oxygen flow ratesAmerican Journal of Veterinary Research, 2006
- The output of an‘empty’ Tec 3 vaporizerAnaesthesia, 1995
- The Effects of Carrier Gas Composition on the Performance of the Tec 6 Desflurane VaporizerAnesthesia & Analgesia, 1994
- Delivery of excess volatile agent due to a vaporizer faultAnaesthesia, 1993
- Long‐term performance of Tec vaporizersAnaesthesia, 1988
- DO MODERN PLENUM VAPORIZERS PROVIDE ACCURATE ANESTHETIC MIXTURES IRRESPECTIVE OF GAS FLOW?Anesthesiology, 1986
- Evaluation of an Isoflurane VaporizerAnesthesia & Analgesia, 1982
- Malfunctioning Halothane VaporizerAnesthesia & Analgesia, 1981
- The Effect of Nitrous Oxide on Halothane Output from Fluotec Mark 2 VaporizersAnesthesiology, 1971
- Malfunction of VaporizersAnesthesiology, 1966