Investigating the Accuracy of the Digihaler, a New Electronic Multidose Dry-Powder Inhaler, in Measuring Inhalation Parameters
Authors:
Open Access
- 1 June 2022
- journal article
- research article
- Published by SAGE Publications in Journal of Aerosol Medicine and Pulmonary Drug Delivery
- Vol. 35 (3) , 166-177
Abstract
Background: The Digihaler ® is a Food and Drug Administration-approved, digital multidose dry powder inhaler with an integrated electronic module that provides patients and health care professionals with feedback on inhalation parameters, including usage, adherence, and technique. This study compared inhalation parameters measured using the Digihaler with readings made simultaneously using an inhalation profile recorder (IPR). Methods: This single-visit, open-label study enrolled children (4–17 years) and adults (18–55 years) with asthma, and adults (≥55 years) with chronic obstructive pulmonary disease (COPD). Participants made three separate inhalations using an empty Digihaler device, each measured simultaneously by the Digihaler and IPR. Inhalation profiles were downloaded from the devices at the end of the study. Inhalation parameters measured included peak inspiratory flow (PIF) and inhaled volume (inhV). The profile with the highest PIF and corresponding IPR profile were analyzed. Results: Overall, 150 participants were enrolled; inhalation data were available for 148 (50 children and 49 adults with asthma, and 49 with COPD). Mean (standard deviation [SD]) age was 39.1 (24.5) years; 51% of participants were male. Overall mean (SD) PIFs as measured by the Digihaler and IPR were 70.62 (17.73) L/min and 72.55 (19.42) L/min, respectively, with a mean percentage difference of −1.75% (95% confidence interval [CI]: −3.64 to 0.15). Mean percentage differences between the Digihaler and IPR measurements of PIF ranged from −2.97% among adults with COPD to 0.16% among children with asthma. Overall mean (SD) inhV for the Digihaler and IPR were 1.57 (0.69) L and 1.67 (0.73) L, respectively, with a mean percentage difference of −6.11 (95% CI: −8.08 to −4.13). There was a strong correlation between PIF and inhV measurements taken by the Digihaler and those taken by the IPR (Spearman's correlation coefficient = 0.96). Conclusions: Our findings confirm the ability of the Digihaler to provide accurate measurement of inhalation parameters when used by patients.Keywords
This publication has 40 references indexed in Scilit:
- Providing Feedback on Adherence Increases Use of Preventive Medication by Asthmatic ChildrenJournal of Asthma, 2010
- Adherence to inhaled therapy, mortality and hospital admission in COPDThorax, 2009
- Can all patients with COPD use the correct inhalation flow with all inhalers and does training help?Respiratory Medicine, 2007
- An audiovisual reminder function improves adherence with inhaled corticosteroid therapy in asthmaJournal of Allergy and Clinical Immunology, 2007
- Patient adherence with COPD therapyEuropean Respiratory Review, 2005
- Adherence to MedicationThe New England journal of medicine, 2005
- Mass Output and Particle Size Distribution of Glucocorticosteroids Emitted from Different Inhalation Devices Depending on Various Inspiratory ParametersJournal of Aerosol Medicine, 2002
- Aerosol deposition in the human lung following administration from a microprocessor controlled pressurised metered dose inhaler.Thorax, 1995
- The Relationship Between Powder Inhaler Resistance and Peak Inspiratory Conditions in Healthy Volunteers — Implications for In Vitro TestingJournal of Aerosol Medicine, 1993
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986