Chromatic Aberration Correction for Atomic Resolution TEM Imaging from 20 to 80 kV
- 9 August 2016
- journal article
- research article
- Published by American Physical Society (APS) in Physical Review Letters
- Vol. 117 (7) , 076101
Abstract
Atomic resolution in transmission electron microscopy of thin and light-atom materials requires a rigorous reduction of the beam energy to reduce knockon damage. However, at the same time, the chromatic aberration deteriorates the resolution of the TEM image dramatically. Within the framework of the SALVE project, we introduce a newly developed corrector that is capable of correcting both the chromatic and the spherical aberration in the range of accelerating voltages from 20 to 80 kV. The corrector allows correcting axial aberrations up to fifth order as well as the dominating off-axial aberrations. Over the entire voltage range, optimum phase-contrast imaging conditions for weak signals from light atoms can be adjusted for an optical aperture of at least 55 mrad. The information transfer within this aperture is no longer limited by chromatic aberrations. We demonstrate the performance of the microscope using the examples of 30 kV phase-contrast TEM images of graphene and molybdenum disulfide, showing unprecedented contrast and resolution that matches image calculations.
Keywords
Funding Information
- Deutsche Forschungsgemeinschaft
- Ministerium für Wissenschaft, Forschung und Kunst Baden-Württemberg
This publication has 29 references indexed in Scilit:
- Accurate Measurement of Electron Beam Induced Displacement Cross Sections for Single-Layer GraphenePhysical Review Letters, 2012
- Chemical Distribution and Bonding of Lithium in Intercalated Graphite: Identification with Optimized Electron Energy Loss SpectroscopyACS Nano, 2011
- Atom-by-atom spectroscopy at graphene edgeNature, 2010
- Atom-by-atom structural and chemical analysis by annular dark-field electron microscopyNature, 2010
- Structural imaging of β-Si3N4 by spherical aberration-corrected high-resolution transmission electron microscopyUltramicroscopy, 2009
- Detection of Single Atoms and Buried Defects in Three Dimensions by Aberration-Corrected Electron Microscope with 0.5-Å Information LimitMicroscopy and Microanalysis, 2008
- Studying Atomic Structures by Aberration-Corrected Transmission Electron MicroscopyScience, 2008
- Radiation damage in the TEM and SEMMicron, 2004
- Irradiation effects in carbon nanostructuresReports on Progress in Physics, 1999