Microfabricated Collector-Generator Electrode Sensor for Measuring Absolute pH and Oxygen Concentrations
- 29 September 2015
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 87 (20) , 10556-10564
Abstract
Fast-scan cyclic voltammetry (FSCV) has attracted attention for studying in vivo neurotransmission due to its subsecond temporal resolution, selectivity, and sensitivity. Traditional FSCV measurements use background subtraction to isolate changes in the local electrochemical environment, providing detailed information on fluctuations in the concentration of electroactive species. This background subtraction removes information about constant or slowly changing concentrations. However, determination of background concentrations is still important for understanding functioning brain tissue. For example, neural activity is known to consume oxygen and produce carbon dioxide which affects local levels of oxygen and pH. Here, we present a microfabricated microelectrode array which uses FSCV to detect the absolute levels of oxygen and pH in vitro. The sensor is a collector-generator electrode array with carbon microelectrodes spaced 5 μm apart. In this work, a periodic potential step is applied at the generator producing transient local changes in the electrochemical environment. The collector electrode continuously performs FSCV enabling these induced changes in concentration to be recorded with the sensitivity and selectivity of FSCV. A negative potential step applied at the generator produces a transient local pH shift at the collector. The generator-induced pH signal is detected using FSCV at the collector and correlated to absolute solution pH by postcalibration of the anodic peak position. In addition, in oxygenated solutions a negative potential step at the generator produces hydrogen peroxide by reducing oxygen. Hydrogen peroxide is detected with FSCV at the collector electrode, and the magnitude of the oxidative peak is proportional to absolute oxygen concentrations. Oxygen interference on the pH signal is minimal and can be accounted for with a postcalibration.Keywords
Funding Information
- U.S. Department of Health and Human Services (DA032530, NS015841)
This publication has 54 references indexed in Scilit:
- Voltammetric pH sensing using carbon electrodes: glassy carbon behaves similarly to EPPGThe Analyst, 2014
- Voltammetric pH sensor based on an edge plane pyrolytic graphite electrodeThe Analyst, 2014
- Fiber-Optic Chemical Sensors and Biosensors (2008–2012)Analytical Chemistry, 2012
- Review Paper: Materials and Techniques forIn VivopH MonitoringIEEE Sensors Journal, 2007
- Brain Tissue Oxygen Concentration MeasurementsAntioxidants & Redox Signaling, 2007
- Spatiotemporal pH dynamics following insertion of neural microelectrode arraysJournal of Neuroscience Methods, 2007
- Fabrication and Use of High-Speed, Concentric H+- and Ca2+-Selective Microelectrodes Suitable for In Vitro Extracellular RecordingJournal of Neurophysiology, 2006
- Anthraquinone-derivatised carbon powder: reagentless voltammetric pH electrodesTalanta, 2003
- International pH Scales and Certification of pHAnalytical Chemistry, 1991
- Depth profile of local oxygen tension and blood flow in rat cerebral cortex, white matter and hippocampusBrain Research, 1988