A Low-Distortion Current-Mode Signal Generator for Wide-Range Bioimpedance Spectroscopy
Authors:
- 21 May 2023
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- p. 1-5
Abstract
This paper presents a low-distortion current-mode sinusoidal signal generator for bioimpedance spectroscopy measurements. The proposed full current-mode operation enables linearity enhancement and potential savings in silicon area and power consumption. Programmability in the low-pass filter and current driver enables impedance measurements from $0.2\ \Omega$ to $10\ \mathrm{k}\Omega$ over a wide frequency range from 1 kHz to 1 MHz. The current generator, designed in a $0.18\ \mu \mathrm{m}$ CMOS process, consumes between $736\ \mu \mathrm{W}$ at the lowest frequency and gain, and 1.70 mW at the highest frequency and gain, and occupies 1.76 mm 2 silicon area. Post-layout simulation results show a spurious-free dynamic range larger than 40 dBc over the entire frequency range, which enables bioimpedance measurements with errors below 1%, as it is required for wearable devices evaluating neuromuscular disorders.
Keywords
This publication has 20 references indexed in Scilit:
- Design of Sub-10-μW Sub-0.1% THD Sinusoidal Current Generator IC for Bio-Impedance SensingIEEE Journal of Solid-State Circuits, 2021
- A 9.6-mW/Ch 10-MHz Wide-Bandwidth Electrical Impedance Tomography IC With Accurate Phase Compensation for Early Breast Cancer DetectionIEEE Journal of Solid-State Circuits, 2020
- A 0.5-V Sub-10-μW 15.28-mΩ/√Hz Bio-Impedance Sensor IC With Sub-1° Phase ErrorIEEE Journal of Solid-State Circuits, 2020
- A Bio-Impedance Readout IC With Digital-Assisted Baseline Cancellation for Two-Electrode MeasurementIEEE Journal of Solid-State Circuits, 2019
- A Multi-Sensor and Parallel Processing SoC for Miniaturized Medical InstrumentationIEEE Journal of Solid-State Circuits, 2018
- A Sinusoidal Signal Generator Using a Constant Gain Finite Impulse Response (FIR) Filter for Electrical Bioimpedance Spectroscopy2018
- Present Uses, Future Applications, and Technical Underpinnings of Electrical Impedance MyographyCurrent Neurology and Neuroscience Reports, 2017
- Electrical Impedance Myography and Its Applications in Neuromuscular DisordersNeurotherapeutics, 2017
- All current-mode frequency selective circuitsElectronics Letters, 1989
- Constant bandwidth voltage amplification using current conveyorsInternational Journal of Electronics, 1988