Postshock Sensing Performance in Transvenous Defibrillation Lead Systems: Analysis of Detection and Redetection of Ventricular Fibrillation
- 1 August 1995
- journal article
- research article
- Published by Wiley in Journal of Cardiovascular Electrophysiology
- Vol. 6 (8) , 604-612
Abstract
The sensing performance of transvenous lead systems may be adversely affected by the delivery of high-energy shocks. This may be due to the proximity of the sensing and energy-delivery electrodes on transvenous leads. The time required for detection of ventricular fibrillation and redetection after a failed first shock was compared in 93 patients with five different lead system-pulse generator combinations: Cadence--Endotak 60 series, Ventak P--Endotak 60 series, Jewel--Transvene, Cadence--TVL, and Cadence--Transvene. A total of 418 successful and 204 failed first shocks were delivered during induced ventricular fibrillation. Redetection times (RED) were consistently shorter than detection times (DET) in the Jewel-Transvene (RED minus DET: -1.9 +/- 0.8 sec, P < 0.0001), the Cadence-TVL (-1.6 +/- 1.0 sec, P < 0.0001), and the Cadence-Transvene combinations (-2.0 +/- 0.9 sec, P < 0.0004). Redetection times were not significantly different than detection times in the Cadence-Endotak combination (0.9 +/- 3.1 sec; P = 0.09). Redetection times were significantly longer than detection times in the Ventak-Endotak combination (1.2 +/- 2.3 sec; P = 0.034). Prolonged individual redetection episodes (> 8.2 sec) were observed in the Cadence-Endotak (7 [10%] of 73 episodes) and the Ventak-Endotak (4 [10%] of 39 episodes), but not in the Jewel-Transvene, the Cadence-TVL, and the Cadence-Transvene combinations. Redetection of ventricular fibrillation may be delayed in some transvenous lead-pulse generator combinations. Successful redetection of ventricular fibrillation following a failed first shock should be demonstrated prior to hospital discharge of patients with implantable defibrillators.Keywords
This publication has 11 references indexed in Scilit:
- Comparison of initial detection and redetection of ventricular fibrillation in a transvenous defibrillator system with automatic gain controlJACC, 1995
- Appropriate sensing of ventricular fibrillation after failed shocks in a transvenous cardioverter-defibrillator system.Circulation, 1994
- ABSTRACT SCSSION 30: THANSVENOUS DEFIBRILLATION LEADSPacing and Clinical Electrophysiology, 1994
- Failure of an implantable cardioverter-defibrillator to redetect ventricular fibrillation in patients with a nonthoracotomy lead system.Circulation, 1992
- Conduction disturbances caused by high current density electric fields.Circulation Research, 1990
- Relation of the intraoperative defibrillation threshold to successful postoperative defibrillation with an automatic implantable cardioverter defibrillatorThe American Journal of Cardiology, 1988
- Muscarinic cholinergic inhibition of adenylate cyclase in airway smooth muscleAmerican Journal of Physiology-Cell Physiology, 1987
- The cellular electrophysiologic changes induced by high-energy electrical ablation in canine myocardium.Circulation, 1986
- Changes in pacing threshold and R wave amplitude after transvenous catheter countershockJACC, 1984
- Pacing threshold changes after transvenous catheter countershockThe American Journal of Cardiology, 1984