Determination of lead in blood using flow injection and a nebuliser interface for flame atomic absorption spectrometry

Authors:
  • ONOlle Nygren
    Olle Nygren
  • CNCarl-Axel Nilsson
    Carl-Axel Nilsson
  • AGAnders Gustavsson
    Anders Gustavsson

Abstract
An improved system for the determination of lead in blood by flame atomic absorption spectrometry has been developed. The system is based on flow injection, a nebuliser interface and a computer signal evaluation system. The system improves the detection limit 12-fold compared with that obtained by unmodified instrumentation. A detection limit of 0.06 µM was obtained using a 10-year-old instrument. With a relative standard deviation of 5% at normal lead levels in blood, even small changes in the lead concentration can be determined with adequate precision and accuracy.