Automatic detection of left ventricular ejection time from a finger photoplethysmographic pulse oximetry waveform: comparison with Doppler aortic measurement
Authors:
- 20 March 2007
- journal article
- research article
- Published by IOP Publishing in Physiological Measurement
- Vol. 28 (4) , 439-452
Abstract
Left ventricular ejection time (LVET) is a useful measure of ventricular performance and preload. The present study explores a novel method of continuous LVET monitoring using a noninvasive finger photoplethysmographic pulse oximetry waveform (PPG-POW). A method for the automatic beat-to-beat detection of LVET from the finger PPG-POW is presented based on a combination of derivative analysis, waveform averaging and rule-based logic. The performance of the detection method was evaluated on 13 healthy subjects during graded head-up tilt. Overall, the correlation between the PPG-POW derived LVET and the aortic flow derived LVET was high and significant (r = 0.897, p < 0.05). The bias was -14 +/- 14 ms (mean +/- SD), and the percentage error was 9.7%. Although these results would not be sufficient to satisfy the requirement for clinical evaluation of LVET when absolute accuracy was demanded, the strong correlation between the PPG-POW LVET and the aortic LVET on an intra-subject basis (r = 0.945 +/- 0.043, mean +/- SD) would support the application of PPG-POW to detect the directional change in LVET of an individual. This could be very useful for the early identification of progressive hypovolaemia or blood loss. The present study has demonstrated a promising approach to extract potentially useful information from a noninvasive, easy-to-obtain signal that could be readily acquired either from existing patient monitoring equipment or from inexpensive instrumentation. More extensive investigation is necessary to evaluate the applicability of the present approach in clinical care monitoring.Keywords
This publication has 24 references indexed in Scilit:
- Testing the Reliability of a New Ultrasonic Cardiac Output Monitor, the USCOM, by Using Aortic Flowprobes in Anesthetized DogsAnesthesia & Analgesia, 2005
- Decrease in Left Ventricular Ejection Time on Digital Arterial Waveform during Simulated Hypovolemia in Normal Humans2004
- Mobile monitoring with wearable photoplethysmographic biosensorsIEEE Engineering in Medicine and Biology Magazine, 2003
- The finger volume pulse and assessment of arterial propertiesJournal Of Hypertension, 2002
- Extracting arterial flow waveforms from pulse oximeter waveformsAnaesthesia, 2001
- Second Derivative of Photoplethysmogram in Children and Young PeopleJapanese Circulation Journal, 2000
- A Meta-Analysis of Studies Using Bias and Precision Statistics to Compare Cardiac Output Measurement TechniquesJournal of Clinical Monitoring and Computing, 1999
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Ascending aortic blood velocity and acceleration using Doppler ultrasound in the assessment of left ventricular functionCardiovascular Research, 1984
- Left ventricular systolic time intervals. A comparison between the conventional carotid pulse curve method and the doppler ultrasound methodEuropean Heart Journal, 1983