A passive membrane system for on‐line mass spectrometry reagent addition
Open Access
- 15 February 2023
- journal article
- research article
- Published by Wiley in Rapid Communications in Mass Spectrometry
- Vol. 37 (8) , e9487
Abstract
Rationale: Post‐separation addition of chemical modifiers in liquid chromatography–mass spectrometry is widely used for improving ionization sensitivity and selectivity. This is typically accomplished using a post‐column T‐junction, which can result in sample dilution and imperfect mixing. We present a passive semi‐permeable hollow fiber membrane approach for the addition of chemical modifiers that avoids these issues. Methods: Model compounds were directly infused by flow injection to an electrospray ionization triple quadrupole mass spectrometer after passing through a polydimethylsiloxane hollow fiber membrane. Ionization enhancement reagents were introduced into the flowing stream by membrane permeation from aqueous solutions. Ionization enhancement from volatile acids and bases in positive and negative electrospray ionization was evaluated to assess the feasibility of this approach. Results: The membrane‐based apparatus resulted in relative ionization enhancement factors of up to 14×, depending upon the analyte, reagent, and ionization mode used. Ionization enhancement signal stability is reasonable (relative standard deviation of 5–7%) for extended periods from the same reagent solution, and minimal analyte dilution is observed. A proof‐of‐concept demonstration of the chromatographic “trifluoroacetic acid fix” strategy is presented. Conclusions: An on‐line mass spectrometry ionization reagent addition method with potential post‐chromatography reagent addition applications was developed using a hollow fiber polydimethylsiloxane membrane. This approach offers a promising alternative to traditional methods requiring additional hardware such as pumps and T‐junctions that can result in sample dilution and imperfect reagent mixing.Keywords
Funding Information
- Natural Sciences and Engineering Research Council of Canada (RGPIN‐2021‐02981, RGPIN‐2022‐05349)
This publication has 32 references indexed in Scilit:
- Membrane-introduction mass spectrometry (MIMS)TrAC Trends in Analytical Chemistry, 2011
- Characterization of a condensed‐phase membrane introduction mass spectrometry (CP‐MIMS) interface using a methanol acceptor phase coupled with electrospray ionization for the continuous on‐line quantitation of polar, low‐volatility analytes at trace levels in complex aqueous samplesRapid Communications in Mass Spectrometry, 2011
- CE‐ESI‐MS analysis of singly charged inorganic and organic anions using a dicationic reagent as a complexing agentElectrophoresis, 2009
- Characterization of Poly(ethylene glycol) and PEGylated Products by LC/MS with Postcolumn Addition of AminesAnalytical Chemistry, 2008
- Enhancing the response of alkyl methylphosphonic acids in negative electrospray ionization liquid chromatography tandem mass spectrometry by post-column addition of organic solventsJournal of the American Society for Mass Spectrometry, 2007
- Simple means to alleviate sensitivity loss by trifluoroacetic acid (TFA) mobile phases in the hydrophilic interaction chromatography–electrospray tandem mass spectrometric (HILIC–ESI/MS/MS) bioanalysis of basic compoundsJournal of Chromatography B, 2005
- On-line identification of sugarcane (Saccharum officinarum L.) methoxyflavones by liquid chromatography–UV detection using post-column derivatization and liquid chromatography–mass spectrometryJournal of Chromatography A, 2005
- Pre‐, on‐ and post‐column derivatization in capillary electrophoresisElectrophoresis, 1997
- LC/MS/MS of carbohydrates with postcolumn addition of metal chlorides using triaxial electrospray probeAnalytical Chemistry, 1995
- Membrane mass spectrometry for the direct trace analysis of volatile organic compounds in air and waterAnalytical Chemistry, 1990