A Nosocomial Outbreak ofLegionella pneumophilaCaused by Contaminated Transesophageal Echocardiography Probes
Authors:
- Nadine TeysseireNTNadine Teysseire
- Jérôme EtienneJEJérôme Etienne
- Didier RaoultDRDidier Raoult
- 1 August 2003
- journal article
- Published by Cambridge University Press (CUP) in Infection Control & Hospital Epidemiology
- Vol. 24 (8) , 619-622
Abstract
A case-control study of three cases ofLegionella pneumophilapneumonia identified transesophageal echocardiography (TEE) as a risk factor. Patient isolates and environmental strains from water used for rinsing TEE probes were identical by pulsed-fleld gel electrophoresis. This is the first report of endoscopy as a potential source of legionellosis.Keywords
This publication has 8 references indexed in Scilit:
- Molecular typing of Legionella pneumophila serogroup 1 isolates from patients and the nosocomial environment by arbitrarily primed PCR and pulsed-field gel electrophoresisJournal of Medical Microbiology, 1999
- Pseudoepidemic of Legionella pneumophila serogroup 6 associated with contaminated bronchoscopesJournal of Hospital Infection, 1997
- Detection of Legionella pneumophila serogroup 1 antigen in nonconcentrated urine and urine concentrated by selective ultrafiltrationJournal of Clinical Microbiology, 1996
- Field inversion gel electrophoretic analysis of Legionella pneumophila strains associated with nosocomial legionellosis in childrenJournal of Clinical Microbiology, 1996
- Interpreting chromosomal DNA restriction patterns produced by pulsed-field gel electrophoresis: criteria for bacterial strain typingJournal of Clinical Microbiology, 1995
- Surveillance for Legionnaires' DiseaseArchives of internal medicine (1960), 1994
- Detection of Legionella spp. in bronchoalveolar lavage fluids by DNA amplificationJournal of Clinical Microbiology, 1992
- Legionnaires' DiseaseNew England Journal of Medicine, 1977