A Study on Ultrasonic Wireless Power Transfer With Phased Array for Biomedical Implants
- 2 June 2023
- journal article
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Biomedical Circuits and Systems
- Vol. 17 (4) , 713-724
Abstract
This paper presents the design, fabrication, and sensitivity analysis of an ultrasound (US) wireless power transfer (WPT) link using an external phased array. Optimal beam focusing and steering is needed for efficient, safe, and reliable US WPT to biomedical implants with millimeter (mm) dimensions. Therefore, the main contributions of this work include the investigation of the 1) performance of the US WPT link using different mm-sized US receivers, 2) effect of different types of errors in the delay profile of the beamforming system on the delivered power, and 3) implant's localization. In measurements, the fabricated 0.94 MHz, 32-element array (39.48×9.6×2 mm 3 ) driven by 25 V pulses with beam focusing and steering capability up to 50 mm depth and ±60° angle could deliver power to different mm-sized US receivers within the FDA safety limit of 720 mW/cm 2 . Specifically, several US transducers with a 1 mm dimension (sphere, cubic, disc shape) and 2 mm dimension (disc shape) received 0.095 mW, 0.25 mW, 0.22 mW, and 0.53 mW, respectively, at a 30 mm depth (0° steering angle). Among these transducers, the sphere shape transducer featured less sensitivity to misalignments. A random error in the phased array delays had a more drastic effect on delivered power reduction. For implant's localization, the measurement results demonstrated comparable power delivery by measuring pulse delays of only 5 elements (out of 32 elements) using 4 different interpolation methods.Keywords
Funding Information
- National Institutes of Health (R21EY030700)
- National Science Foundation (ECCS-1904811)
This publication has 30 references indexed in Scilit:
- Design and Optimization of Ultrasonic Wireless Power Transmission Links for Millimeter-Sized Biomedical ImplantsIEEE Transactions on Biomedical Circuits and Systems, 2016
- Wireless Recording in the Peripheral Nervous System with Ultrasonic Neural DustNeuron, 2016
- Diabetic gastroparesis alters the biomagnetic signature of the gastric slow waveNeurogastroenterology & Motility, 2016
- Ultrasonic beamforming system for interrogating multiple implantable sensors2015
- A Minimally Invasive 64-Channel Wireless μECoG ImplantIEEE Journal of Solid-State Circuits, 2014
- High-resolution Mapping of In Vivo Gastrointestinal Slow Wave Activity Using Flexible Printed Circuit Board Electrodes: Methodology and ValidationAnnals of Biomedical Engineering, 2009
- Gut peristalsis is governed by a multitude of cooperating mechanismsAmerican Journal of Physiology-Gastrointestinal and Liver Physiology, 2009
- Effect of Air Bubbles in the Coupling Medium on Efficacy of Extracorporeal Shock Wave LithotripsyEuropean Urology, 2007
- The Prevalence and Risk Factors of Functional Dyspepsia in a Multiethnic Population in the United StatesAmerican Journal of Gastroenterology, 2004
- Abnormal gastric myoelectrical activity and delayed gastric emptying in patients with symptoms suggestive of gastroparesisDigestive Diseases and Sciences, 1996