Micromachined ultrasound transducer array for cell stimulation with high spatial resolution
- 1 January 2015
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- p. 651-654
Abstract
In this work, we present a piezoelectric micromachined ultrasonic transducer (pMUT) array for local stimulation on cultured cells with a high spatial resolution for the first time. We used a bulk piezoelectric film that has a higher piezoelectric coefficient (d 31 ) than that of the thin film materials to achieve high acoustic power within a small membrane. The 500-μm-wide and 55-μm-thick transducer membrane exhibits a resonant frequency of 780 kHz, which is within the effective frequency range for stimulating cells. We also demonstrate successful in vitro experiments by stimulating cells using the fabricated pMUT array and verifying the stimulation using fluorescence calcium imaging. Using fluorescence imaging, we observed an increase in Ca 2+ level of TRPA1 expressing HEK293T cells under ultrasound sonication, which confirms that TRPA1 channel in HEK293T cells was activated by ultrasound.Keywords
This publication has 14 references indexed in Scilit:
- Fabrication and characterization of cell culture microdevice for nanomechanical stimulation of living cellsAIP Conference Proceedings, 2014
- Ultrasound stimulation increases proliferation of MC3T3-E1 preosteoblast-like cellsJournal of Therapeutic Ultrasound, 2014
- Precise Neural Stimulation in the Retina Using Focused UltrasoundThe Journal of Neuroscience, 2013
- Effective Parameters for Ultrasound-Induced In Vivo NeurostimulationUltrasound in Medicine & Biology, 2013
- A new mechanobiological era: microfluidic pathways to apply and sense forces at the cellular levelCurrent Opinion in Chemical Biology, 2012
- Optogenetic Neuromodulation2012
- Dense type I collagen matrices that support cellular remodeling and microfabrication for studies of tumor angiogenesis and vasculogenesis in vitroBiomaterials, 2010
- Remote Excitation of Neuronal Circuits Using Low-Intensity, Low-Frequency UltrasoundPLOS ONE, 2008
- Noninvasive Human Brain StimulationAnnual Review of Biomedical Engineering, 2007
- TRPA1 is a candidate for the mechanosensitive transduction channel of vertebrate hair cellsNature, 2004