Negligible In Vitro Recovery of Macromolecules from Microdialysis Using 100 kDa Probes and Dextran in Perfusion Fluid
Authors:
Open Access
- 13 March 2024
- journal article
- research article
- Published by Springer Nature in Neurochemical Research
- Vol. 49 (5) , 1322-1330
Abstract
Microdialysis is applied in neurointensive care to monitor cerebral glucose metabolism. If recoverable, macromolecules may also serve as biomarkers in brain disease and provide clues to their passage across the blood–brain barrier. Our study aimed to investigate the in vitro recovery of human micro- and macromolecules using microdialysis catheters and perfusion fluids approved for clinical use. In vitro microdialysis of a bulk solution containing physiological or supraphysiological concentrations of glucose, lactate, pyruvate, human IgG, serum albumin, and hemoglobin was performed using two different catheters and perfusion fluids. One had a membrane cut-off of 20 kDa and was used with a standard CNS perfusion fluid, and the other had a membrane cut-off of 100 kDa and was perfused with the same solution supplemented with dextran. The flow rate was 0.3 µl/min. We used both push and push–pull methods. Dialysate samples were collected at 2-h intervals for 6 h and analyzed for relative recovery of each substance. The mean relative recovery of glucose, pyruvate, and lactate was > 90% in all but two sets of experiments. In contrast, the relative recovery of human IgG, serum albumin, and hemoglobin from both bulk solutions was below the lower limit of quantification (LLOQ). Using a push–pull method, recovery of human IgG, serum albumin, and hemoglobin from a bulk solution with supraphysiological concentrations were above LLOQ but with low relative recovery (range 0.9%–1.6%). In summary, exchanging the microdialysis setup from a 20 kDa catheter with a standard perfusion fluid for a 100 kDa catheter with a perfusion solution containing dextran did not affect the relative recovery of glucose and its metabolites. However, it did not result in any useful recovery of the investigated macromolecules at physiological levels, either with or without a push–pull pump system.Keywords
Funding Information
- National Hospital
This publication has 33 references indexed in Scilit:
- In vivo and in vitro microdialysis sampling of free fatty acidsJournal of Pharmaceutical and Biomedical Analysis, 2007
- Microdialysis for Monitoring Inflammation: Efficient Recovery of Cytokines and Anaphylotoxins Provided Optimal Catheter Pore Size and Fluid Velocity ConditionsScandinavian Journal of Immunology, 2006
- Continuous monitoring of cerebral metabolism in traumatic brain injury: a focus on cerebral microdialysisCurrent Opinion in Critical Care, 2006
- Intracerebral microdialysis in neurosurgical intensive care patients utilising catheters with different molecular cut-off (20 and 100 kD)Acta Neurochirurgica, 2006
- Microdialysis of large moleculesThe AAPS Journal, 2005
- A Microdialysis Technique for Routine Measurement of Macromolecules in the Injured Human BrainNeurosurgery, 2005
- Cerebral microdialysis methodology—evaluation of 20 kDa and 100 kDa cathetersPhysiological Measurement, 2005
- The proteome of human brain microdialysateProteome Science, 2003
- Assaying protein unbound drugs using microdialysis techniquesJournal of Chromatography B, 2003
- Simultaneous monitoring of extracellular glucose, pyruvate, lactate and glutamate in gerbil cortex during focal cerebral ischemia by dual probe microdialysisJournal of Chromatography A, 2001