Comparison of usefulness of computer assisted continuous 48-h 3-lead with 12-lead ECG ischaemia monitoring for detection and quantitation of ischaemia in patients with unstable angina

Authors:

Abstract
The selection of ECG leads used for ST monitoring may influence detection and quantitation of ischaemia. We compared on-line continuous 48-h 12-lead against 3-lead ST monitoring in 130 unstable angina patients (Mortara, ELI-100). Onset and offset of ST episodes were defined by the lead with the first ≥100 μV ST change relative to baseline and the lead with the latest return to baseline ST level, respectively. ST episodes were calculated for 12 leads and 3 leads (V2, V5, III) separately. ST episodes were detected in 88 patients (77%) by 12-lead and in 71 patients (62%) by 3-lead ST monitoring (Ppp<0·0001). The probability of recurrent ischaemia declined most during the first 24 h of monitoring. After a period without ST changes of 1, 12, 24 and 36 h, the probabilities of recurrent ischaemia were 63, 31, 14 and 9%, respectively. Continuous 12-lead ST monitoring increases detection rate and duration of ST episodes compared to 3-lead ST monitoring. The use of continuous 12-lead ECG monitoring devices on emergency wards and coronary care units is recommended.